[en] [en] BACKGROUND: 7T MRI received FDA/CE clearance almost 7 years ago. However, until today, it has not yet been widely adopted in clinical routine. This is mainly due to field inhomogeneities that impede whole-brain coverage. Moreover, the long scan times often associated with high-resolution imaging are an additional limiting factor.
PURPOSE: To combine calibration-free parallel transmit technology (pTx) using universal pulses (UP) with advanced imaging acceleration strategies to achieve homogenous multicontrast 7T MRI with whole-brain coverage and high spatial resolution in short scan time.
MATERIALS AND METHODS: Ten healthy volunteers were scanned both with conventional vendor-provided sequences and with custom sequences for anatomical whole-brain imaging [-weighted, -weighted, FLAIR, and susceptibility-weighted]. The scan times for the 2 anatomical protocols were matched (25 minutes). In addition, a quantitative MRI protocol [multi-parametric mapping (MPM) and chemical exchange saturation transfer (CEST)] was scanned twice using custom sequences with conventional (circular polarized) and UPs, respectively, in a scan time of 2×25 minutes. Moreover, 4 patients with different neurological diseases were scanned, namely temporal lobe epilepsy, spinocerebellar ataxia, cerebral amyloid angiopathy, and glioblastoma. For the patients, only optimized custom sequences with UPs were acquired.
RESULTS: Compared with conventional implementations, the custom sequences provide strongly improved image homogeneity and quality with significantly higher SNR and CNR across the whole brain, including cerebellum and brain stem. Moreover, UPs improve the repeatability of derived quantitative parameters. The suggested protocol has additionally been successfully demonstrated in 4 patients with different neurological pathologies.
CONCLUSIONS: Homogeneous whole-brain 7T MRI with high spatial resolution and high image quality is possible in clinically feasible scan times. The developed protocol can be applied without any expert knowledge and is ready for clinical use. The approach could largely extend applicability of UHF MRI in neuroradiology paving the way for increased routine use of 7T MRI.
Disciplines :
Radiology, nuclear medicine & imaging
Author, co-author :
Faber, Jennifer
Paech, Daniel
Pracht, Eberhard
Stirnberg, Rüdiger
Ehses, Philipp
Völzke, Yannik
Löwen, Daniel
Ferreira, Mónica
Brunheim, Sascha
Boulant, Nicolas
Gras, Vincent
Mauconduit, Franck
Massire, Aurélien
Goa, Pål Erik
Lamalle, Laurent ; Université de Liège - ULiège > Département de physique
This work received financial support through the German Federal Ministry of Education and Research (BMBF; funding code 01ED2109A) and the Research Council of Norway (RCN, funding code 322980), both as part of the SCAIFIELD project under the aegis of the EU Joint Programme\u2014Neurodegenerative Disease Research (JPND) and from the European Union Horizon 2020 Research and Innovation program under grant agreement 885876 (AROMA). D.P. has received funding from the Else Kroener Fresenius Foundation (2022_EKES.33). J.F. was funded within the Advanced Clinician Scientist Programme (ACCENT, funding code 01EO2107, by the German Federal Ministry of Education and Research (BMBF) and as a PI of the iBehave Network, sponsored by the Ministry of Culture and Science of the State of North Rhine-Westphalia and received funding from the National Ataxia Foundation (NAF). J.F. is members of the European Reference Network for Rare Neurological Diseases. T.B. was funded by the Neuro-aCSis Bonn Neuroscience Clinician Scientist Program (2024-12-07).Bundesministerium f\u00FCr Bildung, Wissenschaft, Forschung und Technologie, http://dx.doi.org/10.13039/501100010571, German Federal Ministry of Education and Research (BMBF; funding code 01ED2109A), T.S., Bundesministerium f\u00FCr Bildung und Forschung, http://dx.doi.org/10.13039/501100002347, German Federal Ministry of Education and Research (BMBF; funding code 01ED2109A), T.K., Horizon 2020 Framework Programme, http://dx.doi.org/10.13039/100010661, European Union Horizon 2020 Research and Innovation program under grant agreement 885876 (AROMA)., T.S.
1 Obusez EC Lowe M Oh S-H et al. 7T MR of intracranial pathology: preliminary observations and comparisons to 3T and 1.5T. NeuroImage. 2018;168:459–476.
2 Trattnig S Springer E Bogner W et al. Key clinical benefits of neuroimaging at 7 T. Neuroimage. 2018;168:477–489.
3 Van Lanen RHGJ Colon AJ Wiggins CJ et al. Ultra-high field magnetic resonance imaging in human epilepsy: a systematic review. NeuroImage Clin. 2021;30:102602.
4 Wang I Oh S Blümcke I et al. Value of 7T MRI and post-processing in patients with nonlesional 3T MRI undergoing epilepsy presurgical evaluation. Epilepsia. 2020;61:2509–2520.
5 Platt T Ladd ME Paech D. 7 Tesla and beyond: advanced methods and clinical applications in magnetic resonance imaging. Invest Radiol. 2021;56:705–725.
6 Regnery S Knowles BR Paech D et al. High-resolution FLAIR MRI at 7 Tesla for treatment planning in glioblastoma patients. Radiother Oncol. 2019;130:180–184.
7 Ehses P Brenner D Stirnberg R et al. Whole-brain B1-mapping using three-dimensional DREAM. Magn Reson Med. 2019;82:924–934.
8 Kent JL Dragonu I Valkovič L et al. Rapid 3D absolute B1 + mapping using a sandwiched train presaturated TurboFLASH sequence at 7 T for the brain and heart. Magn Reson Med. 2023;89:964–976.
9 Gras V Vignaud A Amadon A et al. Universal pulses: a new concept for calibration-free parallel transmission. Magn Reson Med. 2017;77:635–643.
10 Van Damme L Mauconduit F Chambrion T et al. Universal nonselective excitation and refocusing pulses with improved robustness to off-resonance for Magnetic Resonance Imaging at 7 Tesla with parallel transmission. Magn Reson Med. 2021;85:678–693.
11 Löwen D Pracht ED Stirnberg R et al. Interleaved binomial k T -points for water-selective imaging at 7T. Magn Reson Med. 2022;88:2564–2572.
12 Gras V Mauconduit F Vignaud A et al. Design of universal parallel-transmit refocusing k T -point pulses and application to 3D T 2 -weighted imaging at 7T. Magn Reson Med. 2018;80:53–65.
13 Gras V Pracht ED Mauconduit F et al. Robust nonadiabatic T2 preparation using universal parallel-transmit kT-point pulses for 3D FLAIR imaging at 7 T. Magn Reson Med. 2019;81:3202–3208.
14 Tse DHY Da Silva NA Poser BA et al. Inhomogeneity mitigation in CEST using parallel transmission. Magn Reson Med. 2017;78:2216–2225.
15 Liebert A Zaiss M Gumbrecht R et al. Multiple interleaved mode saturation (MIMOSA) for B1 + inhomogeneity mitigation in chemical exchange saturation transfer. Magn Reson Med. 2019;82:693–705.
16 Liebert A Tkotz K Herrler J et al. Whole-brain quantitative CEST MRI at 7T using parallel transmission methods and correction. Magn Reson Med. 2021;86:346–362.
17 Völzke Y Akbey S Löwen D et al. Calibration-free whole-brain CEST imaging at 7T with parallel transmit pulse design for saturation homogeneity utilizing universal pulses (PUSHUP). Magn Reson Med. 2025;93:630–642.
18 Leitão D Tomi-Tricot R Bridgen P et al. Parallel transmit pulse design for saturation homogeneity (PUSH) for magnetization transfer imaging at 7T. Magn Reson Med. 2022;88:180–194.
19 Brenner D Stirnberg R Pracht ED et al. Two-dimensional accelerated MP-RAGE imaging with flexible linear reordering. Magma. 2014;27:455–462.
20 Pracht ED Feiweier T Ehses P et al. SAR and scan-time optimized 3D whole-brain double inversion recovery imaging at 7T. Magn Reson Med. 2018;79:2620–2628.
21 Stirnberg R Deistung A Reichenbach JR et al. Rapid submillimeter QSM and R2* mapping using interleaved multishot 3D-EPI at 7 and 3 Tesla. Magn Reson Med. 2024;92:2294–2311.
22 Akbey S Ehses P Stirnberg R et al. Whole-brain snapshot CEST imaging at 7 T using 3D-EPI. Magn Reson Med. 2019;82:1741–1752.
23 Wang D Ehses P Stöcker T et al. Reproducibility of rapid multi-parameter mapping at 3T and 7T with highly segmented and accelerated 3D-EPI. Magn Reson Med. 2022;88:2217–2232.
24 Tabelow K Balteau E Ashburner J et al. hMRI—a toolbox for quantitative MRI in neuroscience and clinical research. Neuroimage. 2019;194:191–210.
25 Zhou J Payen J-F Wilson DA et al. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nat Med. 2003;9:1085–1090.
26 Stöcker T. Recommendations for neuro MRI acquisition strategies. Adv Magn Reson Technol Appl. 2021;4(Elsevier):3–20.
27 Marques JP Kober T Krueger G et al. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. Neuroimage. 2010;49:1271–1281.
28 Charidimou A Boulouis G Frosch MP et al. The Boston criteria version 2.0 for cerebral amyloid angiopathy: A multicentre, retrospective, MRI–neuropathology diagnostic accuracy study. Lancet Neurol. 2022;21:714–725.
29 Vaculčiaková L Podranski K Edwards LJ et al. Combining navigator and optical prospective motion correction for High-quality 500 μ m resolution quantitative Multi-parameter mapping at 7T. Magn Reson Med. 2022;88:787–801.
30 Kraff O Fischer A Nagel AM et al. MRI at 7 tesla and above: demonstrated and potential capabilities. J Magn Reson Imaging. 2014;33:13–33.
31 Cloos M A Boulant N Luong M et al. K(T) -points: Short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume. Magn Reson Med. 2011:67:72–80.
32 Malik SJ Keihaninejad S Hammers A et al. Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses. Magn Reson Med. 2012;67:1303–1315.
33 Herrler J Liebig P Gumbrecht R et al. Fast online-customized (FOCUS) parallel transmission pulses: a combination of universal pulses and individual optimization. Magn Reson Med. 2021;85:3140–3153.
34 Piredda GF Sleight E Yu T et al. Individually optimized dynamic parallel transmit pulses for 3D high-resolution SPACE imaging at 7T. Magn Reson Med. 2025;94:1616–1625.
35 Middlebrooks EH Herrler J Piredda GF et al. Clinical Implementation of Dynamic Parallel Transmission in 7T Brain MRI: Improved Homogeneity and Contrast Using SPACE Sequences. Am J Neuroradiol. 2025;46:2438–2445.
36 Wang ZI Oh S-H Lowe M et al. Radiological and clinical value of 7T MRI for evaluating 3T-visible lesions in pharmacoresistant focal epilepsies. Front Neurol. 2021;12:591586.
37 Opheim G Van Der Kolk A Bloch KM et al. 7T epilepsy task force consensus recommendations on the use of 7T MRI in clinical practice. Neurology. 2021;96:327–341.
38 Ni J Auriel E Martinez-Ramirez S et al. Cortical localization of microbleeds in cerebral amyloid angiopathy: an ultra high-field 7T MRI study. J Alzheimer Dis. 2014;43:1325–1330.
39 Koemans EA Van Etten ES Van Opstal AM et al. Innovative magnetic resonance imaging markers of hereditary cerebral amyloid angiopathy at 7 Tesla. Stroke. 2018;49:1518–1520.
40 Faber J Berger M Wilke C et al. Stage-dependent biomarker changes in spinocerebellar ataxia type 3. Ann Neurol. 2023;95:400–406.
41 Liu Z Zhou X Tao S et al. Application of deep learning accelerated image reconstruction in T2-weighted turbo spin-echo imaging of the brain at 7T. AJNR Am J Neuroradiol. 2025;46:1517–1520.
42 Liu Z Patel V Zhou X et al. Deep-learning reconstruction for 7T MP2RAGE and SPACE MRI: improving image quality at high acceleration factors. AJNR Am J Neuroradiol. 2025;46:2446–2454.
43 Blumenthal M Luo G Schilling M et al. Deep, deep learning with BART. Magn Reson Med. 2023;89:678–693.