Validation, transfer and measurement uncertainty estimation of an HPLC-UV method for the quantification of artemisinin in hydro alcoholic extracts of Artemisia annua L.
ZIME DIAWARA, H.; GBAGUIDI, F.; Evrard, Brigitteet al.
2011 • In Journal of Pharmaceutical and Biomedical Analysis, 56, p. 7-15
[en] Malaria is the world’s most important parasitic infection with 500 millions cases annually and almost 2 millions death per year. This disease is more present in Sub-Saharan Africa where 90% of the infections are found. Artemisinin and its semi synthetic derivatives (artemether, artesunate) have actually the most powerful activity on malaria, even in its complicated forms and resistance cases.
Various methods have been proposed for detection and quantification of artemisinin in Artemisia annua L. by HPLC-UV, but the plant extracts used for this quantification were extracts obtained with organic solvents (toluene, petroleum ether, hexane …). To be able to use crude A. annua extracts prepared at low cost to formulate antipaludic drugs, we chose the use of a mixture of water and ethanol as solvent of extraction, but no adequate analytical method for this kind of extracts is published.
The main objectives of this work were first to develop an analytical method for artemisinin quantification in hydro alcoholic extracts of A. annua. Second, this method had to be thoroughly validated by the research and development laboratory and, third, the transfer of this method to the routine laboratory had to be demonstrated. The final aim was to compare the estimation of measurement uncertainty obtained during the method validation with validation standards to measurement uncertainty estimates obtained during the method transfer study with real samples.
The method was validated following the accuracy profile methodology and was found to be accurate in the concentration range of 10.0 to 54.0 µg/ml with CV <8%. Limit of detection and of quantification were 2.73 and 10.0 µg/ml, respectively. The method was then successfully transferred to a laboratory in Benin by showing that the quality of the results that it will generate during routine application of the method is sufficient. Finally, the measurement uncertainty of the method was estimated from the validation experiments as well as from the transfer study with authentic unspiked samples of A. annua. The comparison of these measurement uncertainty estimations showed that they were coherent. It confirmed thus that the estimation of measurement uncertainty from validation experiments predicts well the measurement uncertainty of real routine samples. This analytical method was thus shown to be convenient for routine analysis of hydro alcoholic extracts of A. annua in Benin.
Disciplines :
Pharmacy, pharmacology & toxicology
Author, co-author :
ZIME DIAWARA, H.
GBAGUIDI, F.
Evrard, Brigitte ; Université de Liège - ULiège > Département de pharmacie > Pharmacie galénique
Quetin-Leclercq, Joëlle
MOUDACHIROU, M.
Debrus, Benjamin ; Université de Liège - ULiège > Département de pharmacie > Chimie analytique
Hubert, Philippe ; Université de Liège - ULiège > Département de pharmacie > Chimie analytique
Rozet, Eric ; Université de Liège - ULiège > Département de pharmacie > Chimie analytique
Language :
English
Title :
Validation, transfer and measurement uncertainty estimation of an HPLC-UV method for the quantification of artemisinin in hydro alcoholic extracts of Artemisia annua L.
La charge du paludisme en Afrique (2003) Le rapport du paludisme en Afrique, pp. 17-27. , OMS OMS/UNICEF, Genève
Klayman, D.L., Qinghaosu (artemisinin): an antimalarial drug from China (1985) Science, 228, pp. 1049-1055
De Vries, P.J., Dien, T.K., Clinical pharmacology and therapeutic potential of artemisinin and its derivates in the treatment of malaria (1996) Drugs, 52, pp. 818-836
Klayman, D.L., Lin, A.J., Acton, N., Scovill, J.P., Hoch, J.M., Milhous, W.K., Theodarides, A.D., Dobek, A.S., Isolation of artemisinin (qinghaosu) from Artemisia annua growing in the United States (1984) J. Nat. Prod., 47, pp. 715-717
Gaudin, M., Simonnet, X., Dosage de l'artémisinine par chromatographie sur couche mince (CCM) (2002) Rev. Suisse Vitic. Arboric. Hortic., 34, pp. 205-208
Zhao, S.S., Zeng, M.Y., Spkectrometrische hochdruck-flüssigkeitschromatographische (HPLC) untersuchungen zur analytik von Quinghaosu (1985) Planta Med., 51, pp. 233-237
Acton, N., Klayman, D.L., Rollman, I.J., Reductive electrochemical HPLC assay for artemisinin (Qinghaosu) (1985) Planta Med., 5, pp. 445-446
Vandenberghe, D.R., Vergauwe, A.N., Montagu, M., Van Den Eeckhout, E.G., Simultaneous determination of artemisinin and its bioprecursors in Artemisia annua (1995) J. Nat. Prod., 58, pp. 798-803
Kohler, M., Werner, H., Christen, P., Veuthey, J.-L., Determination of artemisinin and artemisinic acid by capillary and packed supercritical fluid chromatography (1997) J. High Res. Chromatogr., 20, pp. 62-66
Avery, B.A., Venkatesh, K.K., Avery, M.A., Rapid determination of artemisinin and related analogs by high-performance liquid chromatography and an evaporative light scattering detector (1999) J. Chromatogr. B, 730, pp. 71-80
Bilia, A.R., Melilo de Malgalhaes, P., Bergonzi, M.C., Vincieri, F.F., Simultaneous analysis of artemisinin and flavonoids of several extracts of Artemisia annua L. obtained from a commercial sample and a selected cultivar (2006) Phytomedicine, 13, pp. 487-493
Peng, A.C., Ferreira, J.F.S., Wood, A.J., Direct analysis of artemisinin from Artemisia annua L. using high-performance liquid chromatography with evaporative light scattering detector, and gas chromatography with flame ionization detector (2006) J. Chromatogr. A, 1133, pp. 254-258
Ferreira, J.F.S., Janick, J., Immunoquantitative analysis of artemisinin from Artemisia annua using polyclonal antibodies (1996) Phytochemistry, 41, pp. 97-104
(2006) Monograph on Good Agricultural and Collection Practices (GACP) for Artemisia annua L., , WHO WHO
(2003), 5, pp. 185-233. , The International Pharmocopeia, WHO, Geneva
Sahai, P., Vishwakarma, R.A., Bharel, S., Gulati, A., Abdin, M.Z., Srivastava, P.S., Jain, S.K., HPLC-electrospray ionisation mass spectrometric analysis of antimalarial drug artemisinin (1998) Anal. Chem., 70, pp. 3084-3087
Qian, G.P., Yang, Y.W., Ren, K.L., Determination of artemisinin in Artemisia annua L. by reversed phase HPLC (2005) J. Liq. Chromatogr. Related Technol., 28, pp. 705-712
Christen, P., Veuthen, J.L., Identification and quantification of artemisinin and its derivatives (2001) Curr. Med. Chem., 8, pp. 1827-1839
Marchese, J.A., Rehder, V.L.G., Sartoratto, A., Quantification of artemisinin in Artemisia annua L. - a comparison of thin layer chromatography with densitometric detection and high performance liquid chromatography with UV detection (2001) Rev. Bras. Planta Med., 4, pp. 81-87
Erdemoglu, N., Orhan, L., Kartal, M., Adigüzel, N., Bani, B., Determination of artemisinin in selected artemisia L. species of Turkey by reversed phase HPLC (2007) Rec. Nat. Prod., 1, pp. 36-43
Tonk, S., Bartarya, R., Srivastava, A., Srivastava, S.S., Maharaj Kumari, K., Quantification of artemisinin in different parts of Artemisia annua by HPLC method (2007) Acta Cienc. Indica Chem., 33, pp. 69-74
Zhang, R.S., Wang, Y., Zhao, M., Analysis of artemisinin from Artemisia annua L. by HPLC-UV using pre-column derivatization (2009) Chem. Ind. Forest Prod., 29, pp. 115-119
Rozet, E., Hubert, C., Ceccato, A., Dewé, W., Ziemons, E., Moonen, F., Michail, K., Hubert, P., Using tolerance intervals in pre-study validation of analytical methods to predict in-study results The fit-for-future purpose concept (2007) J. Chromatogr. A, 1158, pp. 126-137. , http://hdl.handle.net/2268/1810
Hubert, P., Nguyen-Huu, J.J., Boulanger, B., Chapuzet, E., Chiap, P., Cohen, N., Compagnon, P.A., valet, L., Harmonization of strategies for the validation of quantitative analytical procedures. A SFSTP proposal? Part I (2004) J. Pharm. Biomed. Anal., 36, pp. 579-586. , http://hdl.handle.net/2268/6169
Dewé, W., Govaerts, B., Boulanger, B., Rozet, E., Chiap, P., Hubert, P., Using total error as decision criterion in analytical method transfer (2007) Chemometr. Intell. Lab. Syst., 85, pp. 262-268. , http://hdl.handle.net/2268/4372
(2001), Guidance for industry: Bioanalytical Method Validation, US Department of Health and Human Services, US Food and Drug Administration (FDA), Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER), Rockville, May
Rozet, E., Ceccato, A., Hubert, C., Ziemons, E., Oprean, R., Rudaz, S., Boulanger, B., Hubert, P., An analysis of recent pharmaceutical regulatory documents on analytical method validation (2007) J. Chromatogr. A, 1158, pp. 111-125. , http://hdl.handle.net/2268/3646
Peters, F.T., Maurer, H.H., Systematic comparison of bias and precision data obtained with multiple-point and one-point calibration in six validated multianalyte assays for quantification of drugs in human plasma (2007) Anal. Chem., 79, pp. 4967-4976
Hubert, P., Nguyen-Huu, J.J., Boulanger, B., Chapuzet, E., Chiap, P., Cohen, N., Compagnon, P.A., Rozet, E., Harmonization of strategies for the validation of quantitative analytical procedures A SFSTP proposal - Part III (2007) J. Pharm. Biomed. Anal., 45, pp. 82-96. , http://hdl.handle.net/2268/6187
Rozet, E., Dewé, W., Ziemons, E., Bouklouze, A., Boulanger, B., Hubert, P., Methodologies for the transfer of analytical methods: a review (2009) J. Chromatogr. B, 877, pp. 2214-2223. , http://hdl.handle.net/2268/18374
(2008), http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_F.pdf, JCGM 100:2008, Evaluation of measurement data - Guide to the expression of uncertainty in measurement (GUM), (accessed 16.03.11)
(2000), Eurachem/Citac Guide Quantifying the Uncertainty in Analytical Measurement, second ed
Feinberg, M., Boulanger, B., Dewe, W., Hubert, P., New advances in chemical data quality: method validation and measurement uncertainty (2004) Anal. Bioanal. Chem., 380, pp. 502-514. , http://hdl.handle.net/2268/6186