MagnResonMed_94_(2025)_134-149_[Fortin]—MPRAGElike--A novel approach to generate T1w images from multi-contrast gradient echo images for brain segmentation.pdf
UHF‐MRI; brain segmentation; multi‐contrast; multi‐parameter mapping; neuroimaging; synthetic MPRAGE; Humans; Male; Adult; Female; Algorithms; Reproducibility of Results; Image Enhancement/methods; Signal-To-Noise Ratio; Sensitivity and Specificity; Brain/diagnostic imaging; Brain/anatomy & histology; Magnetic Resonance Imaging/methods; Image Processing, Computer-Assisted/methods; Image Interpretation, Computer-Assisted/methods; Gradient echo; High SNR; Multiparameters; Scan time; SNR values; Brain; Image Enhancement; Image Interpretation, Computer-Assisted; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Radiology, Nuclear Medicine and Imaging
Abstract :
[en] PURPOSE: Brain segmentation and multi-parameter mapping (MPM) are important steps in neurodegenerative disease characterization. However, acquiring both a high-resolution T1w sequence like MPRAGE (standard input to brain segmentation) and an MPM in the same neuroimaging protocol increases scan time and patient discomfort, making it difficult to combine both in clinical examinations.
METHODS: A novel approach to synthesize T1w images from MPM images, named MPRAGElike, is proposed and compared to the standard technique used to produce synthetic MPRAGE images (synMPRAGE). Twenty-three healthy subjects were scanned with the same imaging protocol at three different 7T sites using universal parallel transmit RF pulses. SNR, CNR, and automatic brain segmentation results from both MPRAGElike and synMPRAGE were compared against an acquired MPRAGE.
RESULTS: The proposed MPRAGElike technique produced higher SNR values than synMPRAGE for all regions evaluated while also having higher CNR values for subcortical structures. MPRAGE was still the image with the highest SNR values overall. For automatic brain segmentation, MPRAGElike outperformed synMPRAGE when compared to MPRAGE (median Dice Similarity Coefficient of 0.90 versus 0.29 and Average Asymmetric Surface Distance of 0.33 versus 2.93 mm, respectively), in addition to being simple, flexible, and considerably more robust to low image quality than synMPRAGE.
CONCLUSION: The MPRAGElike technique can provide a better and more reliable alternative to synMPRAGE as a substitute for MPRAGE, especially when automatic brain segmentation is of interest and scan time is limited.
Disciplines :
Radiology, nuclear medicine & imaging
Author, co-author :
Fortin, Marc-Antoine ; Department of Physics, Norwegian University of Science and Technology, Trondheim, Trøndelag, Norway
Stirnberg, Rüdiger ; German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany
Völzke, Yannik ; German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany
Lamalle, Laurent ; Université de Liège - ULiège > Département de physique
Pracht, Eberhard; German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany
Löwen, Daniel ; German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany
Stöcker, Tony ; German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany ; Department of Physics and Astronomy, University of Bonn, Bonn, Germany
Goa, Pål Erik ; Department of Physics, Norwegian University of Science and Technology, Trondheim, Trøndelag, Norway ; Department of Radiology and Nuclear Medicine, St. Olavs Hospital HF, Trondheim, Norway
Language :
English
Title :
MPRAGElike: A novel approach to generate T1w images from multi-contrast gradient echo images for brain segmentation.
Norges Forskningsråd BMBF - Bundesministerium für Bildung und Forschung F.R.S.-FNRS - Fonds de la Recherche Scientifique TÜBİTAK - Türkiye Bilimsel ve Teknolojik Araştırma Kurumu
Funding text :
This publication is an outcome of SCAIFIELD, an EU Joint Programme\u2013Neurodegenerative Disease Research (JPND) project (https://www.jpnd.eu). The project is supported through the following funding organizations under the aegis of JPND: Belgium, The Fund for Scientific Research (F.R.S.-FNRS; funding code PINT-MULTI/BEJ R.8006.20); Germany, Federal Ministry of Education and Research (BMBF; funding codes 01ED2109A/B); Norway, The Research Council of Norway (RCN; funding code 322980); Turkey, Scientific and Technological Research Council of Turkey (T\u00DCB\u00CDTAK; funding code 121N028).This publication is an outcome of SCAIFIELD, an EU Joint Programme\u2013Neurodegenerative Disease Research (JPND) project ( https://www.jpnd.eu ). The project is supported through the following funding organizations under the aegis of JPND: Belgium, The Fund for Scientific Research (F.R.S.\u2010FNRS; funding code PINT\u2010MULTI/BEJ R.8006.20); Germany, Federal Ministry of Education and Research (BMBF; funding codes 01ED2109A/B); Norway, The Research Council of Norway (RCN; funding code 322980); Turkey, Scientific and Technological Research Council of Turkey (T\u00DCB\u00CDTAK; funding code 121N028).
Mugler III, John P, Brookeman JR. Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE). Magn Reson Med. 1990;15:152-157.
Bruce F. FreeSurfer. Neuroimage. 2012;62:774-781.
Donatas S, Didac V-P, Øystein S, et al. Reliability and sensitivity of two whole-brain segmentation approaches included in FreeSurfer – ASEG and SAMSEG. NeuroImage. 2021;237:118113.
Leonie H, Sailesh C, Santiago E, Kersten D, Bruce F, Martin R. FastSurfer - A fast and accurate deep learning based neuroimaging pipeline. NeuroImage. 2020;219:117012.
Siegfried T, Elisabeth S, Wolfgang B, et al. Key clinical benefits of neuroimaging at 7 T. NeuroImage. 2018;168:477-489.
Fritz S. Whole-body MRI at high field: technical limits and clinical potential. Eur Radiol. 2005;15:946-959.
Webb Andrew G, Collins CM. Parallel transmit and receive technology in high-field magnetic resonance neuroimaging. Int J Imaging Syst Technol. 2010;20:2-13.
Marques José P, Tobias K, Gunnar K, Wietske Z, Pierre-François M, Rolf G. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. NeuroImage. 2010;49:1271-1281.
O'Brien Kieran R, Tobias K, Patric H, et al. Robust T1-weighted structural brain imaging and morphometry at 7T using MP2RAGE. PLoS One. 2014;9:e99676.
Hampus O, Mikael N, Jimmy L, Ronnie W, Gunther H. Radiofrequency bias correction of magnetization prepared rapid gradient echo MRI at 7.0 Tesla using an external reference in a sequential protocol. Tomography. 2021;7:434-451.
Vincent G, Alexandre V, Alexis A, Denis LB, Nicolas B. Universal pulses: A new concept for calibration-free parallel transmission. Magn Reson Med. 2017;77:635-643.
McEvoy LK, Christine F-N, Cooper RJ, et al. Alzheimer Disease: Quantitative Structural Neuroimaging for Detection and Prediction of Clinical and Structural Changes in Mild Cognitive Impairment. Radiology. 2009;251:195-205.
Menke RA, Jan S, Miller KL, et al. MRI characteristics of the substantia nigra in Parkinson's disease: A combined quantitative T1 and DTI study. NeuroImage. 2009;47:435-441.
Gunther H, Henning D, Peter D. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Mag Reson Med. 2008;59:667-672.
Deoni Sean CL, Peters Terry M, Rutt BK. High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2. Mag Reson Med. 2005;53:237-241.
Difei W, Philipp E, Tony S, Rüdiger S. Reproducibility of rapid multi-parameter mapping at 3T and 7T with highly segmented and accelerated 3D-EPI. Magn Reson Med. 2022;88:2217-2232.
Nikolaus W, John S, Guy W, et al. Quantitative multi-parameter mapping of R1, PD*, MT, and R2* at 3T: a multi-center validation. Front Neurosci. 2013;7:46379.
Maryam S, Tobias L, Simon S, et al. Reliability of multi-parameter mapping (MPM) in the cervical cord: A multi-center multi-vendor quantitative MRI study. NeuroImage. 2022;264:119751.
Tobias L, Maryam S, Gunther H, et al. Multiparameter mapping of relaxation (R1, R2*), proton density and magnetization transfer saturation at 3 T: A multicenter dual-vendor reproducibility and repeatability study. Hum Brain Mapp. 2020;41:4232-4247.
Akifumi H, Marcel W, Masaaki H, et al. SyMRI of the brain: rapid quantification of relaxation rates and proton density, with synthetic MRI, automatic brain segmentation, and myelin measurement. Investig Radiol. 2017;52:647.
Ulrike N, Elke H, Oliver B, Julia T, Ralf D. Improved visibility of brain tumors in synthetic MP-RAGE anatomies with pure T1 weighting. NMR Biomed. 2015;28:818-830.
Kanghyun R, Na-Young S, Dong-Hyun K, Yoonho N. Synthesizing T1 weighted MPRAGE image from multi echo GRE images via deep neural network. Magn Reson Imaging. 2019;64:13-20.
Eugenio IJ, Benjamin B, Yaël B, et al. Joint super-resolution and synthesis of 1 mm isotropic MP-RAGE volumes from clinical MRI exams with scans of different orientation, resolution and contrast. NeuroImage. 2021;237:118206.
Simon S, Maryam S, Tobias L, et al. Reliability of spinal cord measures based on synthetic T1-weighted MRI derived from multiparametric mapping (MPM). NeuroImage. 2023;271:120046.
Edwards Luke J, Peter MC, Saskia H, et al. Quantitative MRI maps of human neocortex explored using cell type-specific gene expression analysis. Cereb Cortex. 2023;33:5704-5716.
Rüdiger S, Tony S. Segmented K-space blipped-controlled aliasing in parallel imaging for high spatiotemporal resolution EPI. Magn Reson Med. 2021;85:1540-1551.
Léo VD, Mauconduit F, Thomas C, Nicolas B, Vincent G. Universal nonselective excitation and refocusing pulses with improved robustness to off-resonance for Magnetic Resonance Imaging at 7 Tesla with parallel transmission. Magn Reson Med. 2021;85:678-693.
Breuer Felix A, Martin B, Mueller Matthias F, et al. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA). Magnetic Resonance in Medicine: An Official Journal of the International Society for. Magn Reson Med. 2006;55:549-556.
Yannik V, Suzan A, Daniel L, et al. Calibration-free whole-brain CEST imaging at 7T with parallel transmit pulse design for saturation homogeneity utilizing universal pulses (PUSHUP). Magn Reson Med. 2025;93:630-642.
David L, Raphael T-T, Pip B, et al. Parallel transmit pulse design for saturation homogeneity (PUSH) for magnetization transfer imaging at 7T. Magn Reson Med. 2022;88:180-194.
Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Mag Reson Med. 2007;57:192-200.
Karsten T, Evelyne B, John A, et al. hMRI – A toolbox for quantitative MRI in neuroscience and clinical research. NeuroImage. 2019;194:191-210.
John A. SPM: a history. NeuroImage. 2012;62:791-800.
Gunther H, Henning D, Kai K, Peter D. High-resolution maps of magnetization transfer with inherent correction for RF inhomogeneity and T1 relaxation obtained from 3D FLASH MRI. Mag Reson Med. 2008;60:1396-1407.
Nobuyuki O. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern. 1979;9:62-66.
Zhou W, Bovik Alan C, Sheikh Hamid R, Simoncelli EP. Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Process. 2004;13:600-612.
Tustison Nicholas J, Avants Brian B, Cook Philip A, et al. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging. 2010;29:1310-1320.
Avants Brian B, Nick T, Gang S, et al. Advanced normalization tools (ANTS). Insight J. 2009;2:1-35.
Leonie H, David K, Martin R. FastSurferVINN: Building resolution-independence into deep learning segmentation methods—A solution for HighRes brain MRI. NeuroImage. 2022;251:118933.
Dice LR. Measures of the amount of ecologic association between species. Ecology. 1945;26:297-302.
Thorvald S. A method of establishing groups of equal amplitude in plant sociology based on similarity of species content and its application to analyses of the vegetation on Danish commons. Biologiske Skrifter. 1948;5:1-34.
Tobias H, Bram VG, Styner Martin A, et al. Comparison and evaluation of methods for liver segmentation from CT datasets. IEEE Trans Med Imaging. 2009;28:1251-1265.
Frank W. Individual comparisons by ranking methods. Springer; 1992:196-202.
Carlo B. Teoria statistica delle classi e calcolo delle probabilita. Pubblicazioni del R Istituto Superiore di Scienze Economiche e Commericiali di Firenze. 1936;8:3-62.
Wang Z, Bovik AC. Mean squared error: Love it or leave it? A new look at Signal Fidelity Measures. IEEE Signal Process Mag. 2009;26:98-117.
Marc-Antoine F, Yannik V, Rüdiger S, et al. High quality brain segmentation from multi-parameter mapping at Ultra-High Field MRI. Proceedings of the International Society of Magnetic Resonance in Medicine, 2023, Toronto, Canada; 2023:1776.
Benjamin B, Greve Douglas N, Oula P, et al. SynthSeg: Segmentation of brain MRI scans of any contrast and resolution without retraining. Med Image Anal. 2023;86:102789.
Michele S, Sergio B, Dennis B, Lars M. CEREBRUM-7T: Fast and Fully Volumetric Brain Segmentation of 7 Tesla MR Volumes. Hum Brain Mapp. 2021;42:5563-5580.
Natalia Z, Bruce F, Martin R, Ville R, Polimeni JR. Advantages of cortical surface reconstruction using submillimeter 7 T MEMPRAGE. NeuroImage. 2018;165:11-26.
Louise B, Friedrich Christoph M. Comparison of Automated Volume Extraction with FreeSurfer and FastSurfer for Early Alzheimer's Disease Detection with Machine Learning. 2021 IEEE 34th International Symposium on Computer-Based Medical Systems (CBMS). IEEE; 2021:113-118.
Van Essen DC, Kamil U, Edward A, et al. The Human Connectome Project: a data acquisition perspective. NeuroImage. 2012;62:2222-2231.
Rüdiger S, Yannik V, Daniel L, et al. High-resolution whole-brain multi-parameter mapping at 7 Tesla with interleaved fly-back 3D-EPI and universal pTX pulses. Proceedings of the International Society of Magnetic Resonance in Medicine, 2023, Toronto, Canada; 2023:272.