Device-associated thrombus formation after left atrial appendage occlusion: A systematic review of events reported with the Watchman, the Amplatzer Cardiac Plug and the Amulet.
[en] OBJECTIVES: This study aimed to provide a systematic review of device-associated thrombosis (DAT) after left atrial appendage occlusion (LAAO) with the Watchman, Amplatzer Cardiac Plug, and Amulet devices. BACKGROUND: DAT is known as a complication of LAAO but data about its clinical impact is scarce. METHODS: A systematic review of studies evaluating the incidence, treatment and clinical implications of DAT from January 2008 to September 2015 was conducted. RESULTS: A total of 30 studies describing DAT events were included in the analysis. The overall incidence of DAT was 3.9% (82 DAT for 2118 implanted devices). The median time from procedure to diagnosis of DAT was 1.5 months (IQR: 0-2.9). Most cases were diagnosed with transesophageal echocardiogram (TEE). The treatment consisted of low molecular weight heparin (LMWH) in 45.5% of cases, and oral anticoagulation (OAC) or other treatment modalities in 54.5%. Complete thrombus resolution was achieved in 95.0% of cases (100% with LMWH and 89.5% with OAC). Treatment duration varied greatly with a median treatment duration of 45 days (IQR: 14-135). Clinical events related to DAT consisted of neurologic events namely two transient ischemic attacks (2.4%) and four ischemic strokes (4.9%). CONCLUSIONS: DAT is an infrequent complication of percutaneous LAAO. It occurs mainly early after the procedure and is associated with a low rate of neurological complications. In the majority of cases, diagnosis is made during follow-up imaging with TEE. Anticoagulation treatment seems to be safe and highly effective. Further studies are needed to evaluate the optimal management of DAT. (c) 2017 Wiley Periodicals, Inc.
Disciplines :
Cardiovascular & respiratory systems
Author, co-author :
LEMPEREUR, Mathieu ; Centre Hospitalier Universitaire de Liège - CHU > Département de médecine interne > Cardiologie interventionnelle
Aminian, Adel
Freixa, Xavier
Gafoor, Sameer
Kefer, Joelle
Tzikas, Apostolos
Legrand, Victor ; Université de Liège - ULiège > Département des sciences cliniques > Département des sciences cliniques
Saw, Jacqueline
Language :
English
Title :
Device-associated thrombus formation after left atrial appendage occlusion: A systematic review of events reported with the Watchman, the Amplatzer Cardiac Plug and the Amulet.
Publication date :
2017
Journal title :
Catheterization and Cardiovascular Interventions
ISSN :
1522-1946
eISSN :
1522-726X
Publisher :
John Wiley & Sons, Hoboken, United States - New York
Camm AJ, Lip GYH, De Caterina R, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: An update of the 2010 ESC Guidelines for the management of atrial fibrillation * Developed with the special contribution of the European Heart Rhythm Association. Eur Heart J 2012;33: 2719–2747.
Saw J. Percutaneous left atrial appendage closure for stroke prevention. Trends Cardiovasc Med 2015;26:200–201.
Aminian A, Lalmand J, Tzikas A, et al. Embolization of left atrial appendage closure devices: A systematic review of cases reported with the watchman device and the amplatzer cardiac plug. Catheter Cardiovasc Interv 2015;86:128–135.
Chowdhury MA, Elabbassi W, Al Nooryani A. Left atrial appendage occlusion device-related thrombus: A case report and review of the literature. J Echocardiogr 2012;10:65–66.
Tzikas A, Shakir S, Gafoor S, et al. Left atrial appendage occlusion for stroke prevention in atrial fibrillation: Multicentre experience with the AMPLATZER Cardiac Plug. EuroIntervention 2016;11:1170–1179.
Guerios EE, Schmid M, Gloekler S, et al. Left atrial appendage closure with the Amplatzer cardiac plug in patients with atrial fibrillation. Arq Bras Cardiol 2012;98:528–536.
Lopez-Minguez JR, Eldoayen-Gragera J, Gonzalez-Fernandez R, et al. Immediate and one-year results in 35 consecutive patients after closure of left atrial appendage with the Amplatzer cardiac plug. Rev Esp Cardiol (Engl Ed) 2013;66:90–97.
Plicht B, Konorza TF, Kahlert P, et al. Risk factors for thrombus formation on the Amplatzer Cardiac Plug after left atrial appendage occlusion. JACC Cardiovasc Interv 2013;6:606–613.
Freixa X, Abualsaud A, Chan J, et al. Left atrial appendage occlusion: Initial experience with the Amplatzer Amulet. Int J Cardiol 2014;174:492–496.
Wiebe J, Bertog S, Franke J, et al. Safety of percutaneous left atrial appendage closure with the Amplatzer cardiac plug in patients with atrial fibrillation and contraindications to anticoagulation. Catheter Cardiovasc Interv 2014;83:796–802.
Minguez JR, Asensio JM, Gragera JE, et al. Two-year clinical outcome from the Iberian registry patients after left atrial appendage closure. Heart 2015;101:877–883.
Sick PB, Schuler G, Hauptmann KE, et al. Initial worldwide experience with the WATCHMAN left atrial appendage system for stroke prevention in atrial fibrillation. J Am Coll Cardiol 2007;49:1490–1495.
Bai R, Horton RP, Di Biase L, et al. Intraprocedural and longterm incomplete occlusion of the left atrial appendage following placement of the WATCHMAN device: A single center experience. J Cardiovasc Electrophysiol 2012;23:455–461.
Meincke F, Schmidt-Salzmann M, Kreidel F, et al. New technical and anticoagulation aspects for left atrial appendage closure using the WATCHMAN(R) device in patients not taking warfarin. EuroIntervention 2013;9:463–468.
Ledwoch J, Franke J, Gonzaga M, et al. Left atrial appendage closure: First in man with the 4th generation watchman device. Catheter Cardiovasc Interv 2016;87:787–794.
Chun KR, Bordignon S, Urban V, et al. Left atrial appendage closure followed by 6 weeks of antithrombotic therapy: A prospective single-center experience. Heart Rhythm 2013;10:1792– 1799.
Faustino A, Paiva L, Providencia R, et al. Percutaneous closure of the left atrial appendage for prevention of thromboembolism in atrial fibrillation for patients with contraindication to or failure of oral anticoagulation: A single-center experience. Rev Port Cardiol 2013;32:461–471.
Gafoor S, Franke J, Bertog S, et al. Left atrial appendage occlusion in octogenarians: short-term and 1-year follow-up. Catheter Cardiovasc Interv 2014;83:805.
Matsuo Y, Sandri M, Mangner N, et al. Interventional closure of the left atrial appendage for stroke prevention. Circ J 2014; 78:619–624.
Saw J, Fahmy P, De Jong P, et al. Cardiac CT angiography for device surveillance after endovascular left atrial appendage closure. Eur Heart J Cardiovasc Imaging 2015;16:1198–1206.
Reddy VY, Holmes D, Doshi SK, et al. Safety of percutaneous left atrial appendage closure: results from the Watchman Left Atrial Appendage System for Embolic Protection in Patients with AF (PROTECT AF) clinical trial and the Continued Access Registry. Circulation 2011;123:417–424.
Reddy VY, Mobius-Winkler S, Miller MA, et al. Left atrial appendage closure with the Watchman device in patients with a contraindication for oral anticoagulation: The ASAP study (ASA Plavix Feasibility Study with Watchman Left Atrial Appendage Closure Technology). J Am Coll Cardiol 2013;61: 2551–2556.
Lammers J, Elenbaas T, Meijer A. Thrombus formation on an Amplatzer closure device after left atrial appendage closure. Eur Heart J 2013;34:741.
Dilling-Boer D, Benit E, et al. Late organized left atrial thrombus on a left atrial appendage closure device. J Cardiovasc Electrophysiol 2014;25:445–446.
Schroeter MR, Danner BC, Hunlich M, et al. Uncommon delayed and late complications after percutaneous left atrial appendage closure with Amplatzer((R)) Cardiac Plug. Clin Res Cardiol 2014;103:285–290.
Ternacle J, Lellouche N, Deux JF, et al. Left atrial appendage occluder thrombosis after successful implantation. Circulation 2014;129:2576–2577.
Freixa X, Scalone G, Martin-Yuste V, et al. Large protruding thrombus over left atrial appendage occlusion device successfully treated with apixaban. Eur Heart J 2015;36:1427.
Schiettekatte S, Czapla J, Nijs J, et al. Unmasking a naked left atrial appendage closure device: A case of a silent embolic threat. Heart Rhythm 2014;11:2314–2315.
Canavy I, Colin R, Desfossez L, et al. [Detection and surveillance of left atrial thrombosis by transesophageal echocardiography]. Arch Mal Coeur Vaiss 1999;92:35–42.
Gasparini M, Ceriotti C, Bragato R. Huge left atrial thrombus after left atrial appendage occlusion with a Watchman device. Eur Heart J 2012;33:1998.
Elabbassi W, Osman AN, Chowdhury MA, et al. Device-related thrombus overlying a left-atrial occlusion device implanted for persistent spontaneous echo contrast despite prolonged oral anticoagulation therapy. Cardiology 2013;125:78–81.
Kar S, Hou D, Jones R, et al. Impact of Watchman and Amplatzer devices on left atrial appendage adjacent structures and healing response in a canine model. JACC Cardiovasc Interv 2014; 7:801–809.
Holmes DR, Jr., Kar S, Price MJ, et al. Prospective randomized evaluation of the Watchman left atrial appendage closure device in patients with atrial fibrillation versus long-term warfarin therapy: The PREVAIL trial. J Am Coll Cardiol 2014;64:1–12.
Rosenzweig BP, Katz E, Kort S, et al.Thromboembolus from a ligated left atrial appendage. J Am Soc Echocardiogr 2001;14: 396–398.
Fernandez-Rodriguez D, Vannini L, Martin-Yuste V, et al. Medical management of connector pin thrombosis with the Amplatzer cardiac plug left atrial closure device. World J Cardiol 2013;5:391–393.
Lam SC, Bertog S, Sievert H. Incomplete left atrial appendage occlusion and thrombus formation after Watchman implantation treated with anticoagulation followed by further transcatheter closure with a second-generation Amplatzer Cardiac Plug (Amulet device). Catheter Cardiovasc Interv 2015;85:321–327.
Schwartz RS, Holmes DR, Van Tassel RA, et al. Left atrial appendage obliteration: mechanisms of healing and intracardiac integration. JACC Cardiovasc Interv 2010;3:870–877.
Holmes DR, Reddy VY, Turi ZG, et al. Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: A randomised noninferiority trial. Lancet 2009;374:534–542.
Rodes-Cabau J, Champagne J, Bernier M. Transcatheter closure of the left atrial appendage: Initial experience with the Amplatzer cardiac plug device. Catheter Cardiovasc Interv 2010;76:186–192.
Ruiz-Garcia J, Moreno R. Percutaneous closure of left atrial appendage: Device-indicated antiplatelet therapy may also lead to fatal bleeding. A call for evidence-based antiplatelet regimen. J Thromb Thrombolysis 2014;37:359–361.
Freixa X, Chan JL, Tzikas A, et al. The Amplatzer Cardiac Plug 2 for left atrial appendage occlusion: Novel features and first-in-man experience. EuroIntervention 2013;8:1094–1098.
Wysokinski WE, Ammash N, Sobande F, et al. Predicting left atrial thrombi in atrial fibrillation. Am Heart J 2010;159:665– 671.
Schneider B, Stollberger C, Sievers HH. Surgical closure of the left atrial appendage—A beneficial procedure? Cardiology 2005; 104:127–132.