Article (Scientific journals)
A pilot study of the safety and performance of the Otologics fully implantable hearing device: transducing sounds via the round window membrane to the inner ear.
Lefèbvre, Philippe; Martin, Christine; Dubreuil, C. et al.
2009In Audiology and Neuro-otology, 14 (3), p. 172-80
Peer Reviewed verified by ORBi
 

Files


Full Text
AUdio_neurooto_Lefebvre_2009.pdf
Publisher postprint (417.79 kB)
Request a copy

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Acoustic Stimulation; Audiometry, Pure-Tone; Auditory Threshold; Evoked Potentials, Auditory; Follow-Up Studies; Hearing Loss, Conductive/physiopathology/surgery; Hearing Loss, Sensorineural/physiopathology/surgery; Humans; Monitoring, Intraoperative; Pilot Projects; Prosthesis Implantation/methods; Round Window, Ear/physiopathology/surgery; Safety; Speech Perception
Abstract :
[en] OBJECTIVES: The safety and performance of the Otologics fully implantable hearing device were assessed in adult patients with mixed conductive and sensorineural hearing loss. METHODS: The subcutaneous microphone of this fully implantable device picks up ambient sounds, converts them into an electrical signal, amplifies the signal according to the user's needs, and sends it to an electromechanical transducer. The transducer tip is customized with a prosthesis in order to be in contact with the round window membrane and is protected by fascia; this translates the electrical signal into a mechanical motion that directly stimulates the round window membrane and enables the user to perceive sound. The implanted battery is recharged daily via an external charger and the user can turn the implant on and off as well as adjust the volume with a hand-held remote control. In this pilot study, 6 patients with mixed conductive and sensorineural hearing loss were implanted with the Otologics fully implantable hearing device. Pre- and postoperative air conduction, bone conduction, as well as aided and unaided thresholds and speech scores were measured. RESULTS: No significant differences between preoperative and postoperative pure-tone averages were noted. Average improvement ranged from 19.16 to 35.8 dB of functional gain across audiometric frequencies with a mean of 26.17 +/- 5.15 dB. Long-term average functional gain at 12 months was 20.83 +/- 6.22 dB. Word recognition scores demonstrated significant differences between unaided and implant-aided conditions. CONCLUSIONS: Preliminary results of this trial of the Otologics fully implantable hearing device provide evidence that this fully implantable device is capable of efficiently transferring the sound to the inner ear via the round window membrane in patients with mixed hearing loss.
Disciplines :
Otolaryngology
Gastroenterology & hepatology
Author, co-author :
Lefèbvre, Philippe ;  Université de Liège - ULiège > Département des sciences cliniques > Oto-rhino-laryngologie et audiophonologie
Martin, Christine ;  Centre Hospitalier Universitaire de Liège - CHU > HOSPITALISATION - GASTRO & NEPHROLOGIE (SN +4B)
Dubreuil, C.
Decat, M.
Yazbeck, A.
Kasic, J.
Tringali, S.
Language :
English
Title :
A pilot study of the safety and performance of the Otologics fully implantable hearing device: transducing sounds via the round window membrane to the inner ear.
Publication date :
2009
Journal title :
Audiology and Neuro-otology
ISSN :
1420-3030
eISSN :
1421-9700
Publisher :
S. Karger, Basel, Switzerland
Volume :
14
Issue :
3
Pages :
172-80
Peer reviewed :
Peer Reviewed verified by ORBi
Commentary :
Copyright 2008 S. Karger AG, Basel.
Available on ORBi :
since 19 March 2010

Statistics


Number of views
167 (4 by ULiège)
Number of downloads
2 (0 by ULiège)

Scopus citations®
 
36
Scopus citations®
without self-citations
29
OpenCitations
 
33
OpenAlex citations
 
41

Bibliography


Similar publications



Contact ORBi