Article (Scientific journals)
Report of an incident during the use of the Servo-ventilator 900 in paediatric anesthesia.
Degauque, C.; Philippart, C.; Lamy, Maurice
1977In Acta Anaesthesiologica Belgica, 28 (3), p. 199-207
Peer Reviewed verified by ORBi
 

Files


Full Text
55.pdf
Publisher postprint (485.24 kB)
Request a copy

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Anesthesia, General/adverse effects; Humans; Infant; Male; Respiration, Artificial/adverse effects; Ventilators, Mechanical
Abstract :
[en] The authors report the occurrence of an incident during the use of the Servo-ventilator 900 for paediatric anesthesia; it consisted in self-oscillation of the gas column contained in the patient respiratory circuit. After having analysed the incident and outlined the various ways of avoiding it, attention is drawn to the importance of the total circuit volume of the respirator. The respiratory circuit chosen for ventilating children, has a compliance of 1.3 ml/cm H2O.
Disciplines :
Anesthesia & intensive care
Author, co-author :
Degauque, C.
Philippart, C.
Lamy, Maurice ;  Université de Liège - ULiège > Département des sciences cliniques > Département des sciences cliniques
Language :
English
Title :
Report of an incident during the use of the Servo-ventilator 900 in paediatric anesthesia.
Publication date :
1977
Journal title :
Acta Anaesthesiologica Belgica
ISSN :
0001-5164
eISSN :
2736-5239
Publisher :
Acta Medica Belgica, Bruxelles, Belgium
Volume :
28
Issue :
3
Pages :
199-207
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 28 May 2009

Statistics


Number of views
37 (0 by ULiège)
Number of downloads
0 (0 by ULiège)

Scopus citations®
 
0
Scopus citations®
without self-citations
0

Bibliography


Similar publications



Contact ORBi