Abstract :
[en] Objectives: Validation of the Architect 25-OH vitamin D assay.
Design and methods: Determination of repeatability, reproducibility, accuracy profile and 25(OH)-vitamin D2 recovery on native samples. Comparison with DiaSorin Liaison and RIA.
Results and conclusion: Coefficients of variation: b6% (13.6 ng/mL) and 2.2% (78.1 ng/mL). Functional sensitivity: 5 ng/mL. Accuracy profile shows that the method is validated between 13.6 and 78.1 ng/mL. Recovery of 25(OH)D2: 75,8%( 95% CI: 61.9–89.7%). Good correlation with DiaSorin RIA and Liaison b50 ng/mL; above this threshold a systematic positive bias was observed.
Scopus citations®
without self-citations
21