[en] This study compared performance of three heat and moisture exchangers (HME) during short-term postoperative mechanical ventilation. Temperature and absolute humidity (AH) were measured at various points of the ventilatory circuit. There was no statistical difference between the groups, regarding ambient and body To, body weight, fraction of inspired oxygen, tidal volume, and respiratory rate. Only the hygroscopic HME (groups 2 and 3) provide adequate conditioning with regard to AH and To of the inspiratory gases. The performance of hydrophobic HME (group 1) was inferior and appears to be unsatisfactory. Indirect evaluation (variations of inspiratory gases and tracheal temperatures, AH of the expired gases) confirmed the superiority of the hygroscopic HME. These data suggest that humidification of inspiratory gases with a hygroscopic HME is a defensible practice during short-term postoperative mechanical ventilation. Performance of hydrophobic HME may be weak and can expose the patient to an unacceptable risk of endotracheal tube occlusion.
Disciplines :
Anesthesia & intensive care
Author, co-author :
Sottiaux, Thierry
Mignolet, Ghislaine ; Centre Hospitalier Universitaire de Liège - CHU > Médecine de l'appareil locomoteur
Damas, Pierre ; Université de Liège - ULiège > Soins intensifs
Lamy, Maurice ; Université de Liège - ULiège > Département des sciences cliniques > Anesthésie et réanimation
Language :
English
Title :
Comparative evaluation of three heat and moisture exchangers during short-term postoperative mechanical ventilation
J Chalon C Patel M Ali S Ramanathan L Capan CK Tang Humidity and the anesthetized patient Anesthesiology 50 1979 195 198
J Chalon Low humidity and damage to tracheal mucosa Bull NY Acad Med 56 1980 314 322
EW Fonkalrud M Sanchez I Higashijima E Arima A comparative study of the effects of dry vs humidified ventilation of canine lungs Surgery 78 1975 373 380
AR Forbes Humidification and mucus flow in the intubated trachea Br J Anesth 45 1973 874 878
JA Hirsch JL Tokayer MJ Robinson MA Sackner Effects of dry air and subsequent humidification on tracheal mucous velocity in dogs J Appl Physiol 39 1975 242 246
U Mercke The influence of varying air humidity on mucociliary activity Acta Otolaryngol 79 1975 133 139
K Rashad K Wilson HH Hurt Effects of humidification of anesthetic gases on static compliance Anesth Analg 46 1967 126 133
HL Cadwallader CR Bradley GAJ Ayliffe Bacterial contamination and frequency of changing ventilator circuitry J Hospital Infection 15 1990 65 72
DE Craven MG Connolly DA Lichtenberg PJ Primeau WR McCabe Contamination of mechanical ventilators with tubing changes every 24 hours N Engl J Med 25 1982 1505 1509
DE Craven TA Gourlate BJ Make Contaminated condensate in mechanical ventilator circuits: risk factor for nosocomial pneumonia Am Rev Respir Dis 129 1984 625 628
D Dreyfuss K Djedaini P Weber P Brun JJ Lanore J Rahmani Prospective study of nosocomial pneumonia and of patient and circuit colonization during mechanical ventilation with circuit changes every 48 hours versus no change Am Rev Respir Dis 143 1991 738 743
NG Torelam A heat-and-moisture exchanger for post-tracheotomy care Acta Otolaryngol 52 1960 461 471
J Chalon JP Markham M Ali S Ramanathan H Turndorf The Pall Ultipor breathing circuit filter: an efficient heat and moisture exchangers Anesth Analg 63 1984 566 570
M Cigada A Elena G Solca G Damia The efficiency of twelve heat and moisture exchangers: an in vitro evaluation Int Care World 2 1990 98 101
A Gedeon C Mebius The hygroscopic condenser humidifier: a new device for general use in anesthesia and intensive care Anesthesia 34 1979 1043 1047
A Gedeon C Mebius K Palmer Neonatal hygroscopic condenser humidifier Crit Care Med 15 1987 51 54
C Mebius A comparative evaluation of disposable humidifiers Acta Anesthesiol Scand 27 1983 403 409
M Ogino R Kopotic FL Mannino Moisture-conserving efficiency of condenser humidifiers Anesthesia 40 1985 990 995
FP Primiano ME Moranz FW Montague RB Miller DP Sachs Conditioning of inspired air by hygroscopic condenser humidifier Crit Care Med 12 1984 675 678
MP Shelly GM Lloyd GR Park A review of the mechanisms and methods of humidification of inspired gases Int Care Med 14 1988 1 9
DJ Steward A disposable condenser humidifier for use during anesthesia Canad Anesth Soc J 23 1976 191 195
AK Walker DW Bethune A comparative study of condenser humidifiers Anaesthesia 3 1976 1086 1093
DB Weeks FM Ramsey Laboratory investigation of six artificial noses for use during endotracheal anesthesia Anesth Analg 62 1983 758 763
NR Mac Intyre HR Anderson RM Silver FR Schuler RE Coleman Pulmonary function in mechanically-ventilated patients during 24 hours use of a hygroscopic condensor humidifier Chest 84 1983 561 564
A Tenaillon G Cholley R Boiteau D Perrin-Gachadoat M Burdin Filtre échangeur de chaleur et d'humidité versus humidificateur chauffant en ventilation mécanique prolonée Réan Soins Intens Med Urg 5 1989 5 10
M Bouilhac Filtre bactérien pour circuit de ventilation Agressologie 25 1984 299 302
S Bygdeman C Von Euler B Nystrom Moisture exchangers do not prevent patient contamination of ventilators: a microbiological study Acta Anesthesiol Scand 28 1984 591 594
J Gallagher JE Strangeways J Allt-Graham Contamination control in long-term ventilation Anesthesia 42 1987 476 481
DJ Powner CS Sanders BJ Bailey Bacteriologic evaluation of the Servo 150 hygroscopic condenser-humidifier Crit Care Med 14 1986 135 137
LD Saravolatz DJ Pohold W Conway W Haberaecker NP Markowitz J Popovich Lack of bacterial aerosols associated with heat and moisture exchangers Am Rev Respir Dis 134 1986 21416
F Squadrini V Borghi A Sabbattini F Taparelli N Mongiardo Efficacité des filtres aux tests pour l'élimination des virus Filtration, humidification et réchauffement en ventilation mécanique. 18e Congrès de la Société de Réanimation de Langue Française, Paris 1989
K Stange S Bygdeman Do moisture exchangers prevent patient contamination of ventilators? Acta Anesthesiol Scand 24 1980 487 490
R Hay WC Miller Efficacy of a new hygroscopic condenser humidifier Crit Care Med 10 1982 49 51
IL Cohen PF Weinberg A Fein GS Rowinski Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit Crit Care Med 16 1988 277 279
C Martin G Perrin MJ Gevaudan P Saux F Gouin Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit Chest 97 1990 144 149
SA Persch AM Realey Effectiveness of the Servo SH 150 “Artificial Nose” humidifier: a case report Resp Care 29 1984 1009 1012
C Martin L Papazian G Perrin A Fraisse F Gouin Humidification chauffants ou échangeurs de chaleur et d'humidité lors de ventilation <10L/MIN [abstract] Réanimation Urgences 1 1992 68
B Misset B Escudier D Rivara B Leclercq G Nitenberg Heat and moisture exchanger vs heated humidifier during long-term mechanical ventilation: a prospective randomized study Chest 100 1991 160 163