Evaluation of frequency and type of errors detected by a computerized record and verify system during radiation treatment
Barthelemy, Nicole; Sabatier, Jacques; Dewé, Walthère et al.
1999In Radiotherapy and Oncology, 53 (2), p. 149-54
Peer Reviewed verified by ORBi
 

Files


Full Text
Evaluation of frequencity and type of errors detected by a computerized record and verify system during radiation treatment.pdf
Publisher postprint (112.5 kB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Abstract :
[en] Background: Computerized record and verify systems (RVS) have been introduced to improve the precision of radiation treatment delivery. These systems prevent the delivery of ionizing radiations when the settings of the treatment machine do not match the intended parameters within some maximal authorized deviation. Purpose: To assess the potential alteration of the frequency of errors associated with the use of RVS during radiation treatment delivery. Materials and methods: The software of the RVS was altered in order to record the settings actually used for radiation treatment delivery whereas the verification function was suppressed. At the end of the study period, the settings used during daily administration of radiation treatment were compared to the parameters recorded in the RVS using the computer. They were also compared with the planned ones written in the patient treatment chart. Results: Out of the 147 476 parameters examined during the study period, 678 (0.46%) were set erroneously. At least one error occurred in 628 (3.22%) of the 19 512 treated fields. An erroneous parameter was introduced in the RVS memory in 22 (1.17%) of the 1885 fields. Conclusions: RVS has the potential to improve precision of radiation treatment delivery by detecting a significant number of setting errors. However, excessive confidence in RVS could lead to repeated errors as there is a potential for the entry of erroneous parameters into the RVS memory.
Disciplines :
Oncology
Author, co-author :
Barthelemy, Nicole ;  Université de Liège - ULiège > Radiothérapie
Sabatier, Jacques;  Centre Hospitalier Universitaire de Liège - CHU > Radiotherapy
Dewé, Walthère ;  Centre Hospitalier Universitaire de Liège - CHU > Medical Informatics and Statistics
Albert, Adelin  ;  Centre Hospitalier Universitaire de Liège - CHU > Medical Informatics and Statistics
Deneufbourg, Jean-Marie  ;  Centre Hospitalier Universitaire de Liège - CHU > Radiotherapy
Language :
English
Title :
Evaluation of frequency and type of errors detected by a computerized record and verify system during radiation treatment
Publication date :
November 1999
Journal title :
Radiotherapy and Oncology
ISSN :
0167-8140
eISSN :
1879-0887
Publisher :
Elsevier Scientific, Limerick, Ireland
Volume :
53
Issue :
2
Pages :
149-54
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 19 January 2010

Statistics


Number of views
103 (11 by ULiège)
Number of downloads
376 (0 by ULiège)

Scopus citations®
 
52
Scopus citations®
without self-citations
52
OpenCitations
 
42

Bibliography


Similar publications



Contact ORBi