[en] Surfactant-based skin care products can induce minimal to more severe alterations to human stratum corneum (SC). These changes can be conveniently assessed using electrometric devices. Four different surfactant-based solutions were applied on the forearms of 10 healthy volunteers following a soap chamber test method. The 2-day patch testing was followed by a 2-week follow-up. Clinical and instrumental assessments were performed at 7 time points. Data generated by the Corneometer® CM820, the DermaLab Moisture Module® and the Nova Dermal Phase Meter (DPM)® 9003 were compared. The significant differences between the product aggressiveness to the SC were similarly disclosed by the Corneometer® and the DermaLab Moisture Module®. The ranking of product aggressiveness given by the 3 devices 5 days after patch removal was similar to that given by the clinical assessment of erythema. A high level of correlation were found between the values yielded by the immediate and continuous mode assessments using the DermaLab Moisture Module with the Pin Probe, while the correlation was weaker with the Nova DPM®.
This finding may be related to the fact that the Nova DPM is influenced by the accumulation of the transepidermal water loss in the continuous mode assessment while there is no significant occlusion effect of the DermaLab Moisture® Pin Probe that is more representative of the initial level of SC hydration. In conclusion, the Corneometer® and the DermaLab Moisture Module® respond similarly to changes in the SC content. This finding is different from some previous reports.
Disciplines :
Dermatology
Author, co-author :
Uhoda, Emmanuelle ; Centre Hospitalier Universitaire de Liège - CHU > Dermatopathologie
Paye, Marc; Colgate Palmolive, Herstal, Belgium
Pierard, Gérald ; Centre Hospitalier Universitaire de Liège - CHU > Dermatopathologie
Language :
English
Title :
Comparative clinical and electrometric assessments of the impact of surfactants on forearm skin.
Polefka T: Surfactant interactions with skin; in Broze G (ed): Handbook of Detergents. New York, Dekker, 1999, part A: Properties, pp 443-468.
Piérard G-E, Goffin V, Hermanns-Lê T, Arrese JE, Piérard-Franchimont C: Surfactant induced dermatitis: A comparison of corneosurfametry with predictive testing on human and reconstructed skin. J Am Acad Dermatol 1995;33:462-469.
Zhai H, Fautz R, Fuchs A, Branco N, Maibach HI: Assessment of the subclinical irritation of surfactants: A screening open assay model. Exog Dermatol 2002;1:238-241.
Agner T, Serup J: Sodium lauryl sulfate for irritant patch testing: A dose-response study using bioengineering methods for determination of skin irritation. J Invest Dermatol 1990;95:543-547.
Widmer J, Elsner P, Burg G: Skin irritant reactivity following experimental cumulative irritant contact dermatitis. Contact Dermatitis 1994;30:35-39.
Paye M, Simion A, Piérard GE: Dansyl chloride labelling of stratum corneum: Its rapid extraction from skin can predict skin irritation due to surfactants and cleansing products. Contact Dermatitis 1994;30:91-96.
Frosch PJ, Kligman AM: The soap chamber test: A new method for assessing irritancy of soaps. J Am Acad Dermatol 1979;1:35-41.
Simion AF, Rhein LD: Sequential order of skin response to surfactants during soap chamber test. Contact Dermatitis 1991;25:242-249.
Serup J: EEMCO guidance for the assessment of dry skin (xerosis) and ichthyosis: Clinical scoring systems. Skin Res Technol 1995;1:109-114.
Berardesca E: EEMCO guidance for the assessment of stratum corneum hydration: Electrical methods. Skin Res Technol 1997;3:126-132.
Treffel P, Gabard B: Stratum corneum dynamic function measurements after moisturizer or irritant application. Arch Dermatol Res 1995;287:474-479.
Paquet F, Piérard-Franchimont C, Fumal I, Goffin V, Paye M, Piérard GE: Sensitive skin at menopause: Dew point and electrometric properties of the stratum corneum. Maturitas 1998;28:221-227.
Claes G, Piérard GE: Biometrological assessment of skin protectors against moderate cold threat. Exog Dermatol 2002;1:92-96.
Paye M, Van de Gaer D, Morrison BM Jr: Corneometry measurements to evaluate skin dryness in the modified soap chamber test. Skin Res Technol 1995;1:123-127.
Fluhr JW, Gloor V: Comparative study of five instruments measuring stratum corneum hydration. Skin Res Technol 1999;5:171-178.
Morrison MB, Scala D: Comparison of instrumental measurements of skin hydration. J Toxicol Cut Ocular Toxicol 1996;15:305-314.
Loden M, Hagforsen E, Lindberg M: The presence of body hair influences the measurements of skin hydration with the Corneometer. Acta Derm Venereol 1995;75:449-450.
Goffin V, Piérard-Franchimont C, Piérard GE: Passive sustainable hydration of the stratum corneum following surfactant challenge. Clin Exp Dermatol 1999;24:308-311.