Development and validation by accuracy profile of a method for the analysis of monoterpenes in indoor air by active sampling and thermal desorption-gas chromatography-mass spectrometry
Marlet, Christelle; Lognay, Georges
2010In Talanta, 82 (4), p. 1230-1239
Peer Reviewed verified by ORBi
 

Files


Full Text
TAL11395_PostPrintAuteur.pdf
Author postprint (1.2 MB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Monoterpenes; Accuracy profile; Thermal desorption; Air sampling; Tenax TA; Wood construction
Abstract :
[en] The technique of thermal desorption (TD)-GC/MS was evaluated for the measurement of monoterpenes in indoor air. The validation strategy was intentionally oriented towards routine use and the reliability of the method rather than extreme performance. For this reason, validation by accuracy profile was chosen. The accuracy profile procedure, which is based on the concept of total error (bias + standard deviation), guarantees that a known proportion of future results obtained with the method will be within acceptance limits. For all the compounds tested in the present study, α-pinene, α-terpineol, β-pinene, d-limonene, Δ3-carene, camphene, 1,8-cineole, p-cymene, linalool, but not in the case of carvone, the accuracy profile procedure established that at least 95% of the future results obtained would be within the ±15% acceptance limits of the validated method over the whole defined concentration range. Other parameters, such as selectivity, recovery, repeatability, stability of the molecules of interest and the effect of temperature, were also determined. The performance of the described method was finally evaluated by the analysis of indoor air from new timber frame constructions.
Disciplines :
Environmental sciences & ecology
Chemistry
Author, co-author :
Marlet, Christelle ;  Université de Liège - ULiège > Chimie et bio-industries > Chimie analytique
Lognay, Georges ;  Université de Liège - ULiège > Chimie et bio-industries > Chimie analytique
Language :
English
Title :
Development and validation by accuracy profile of a method for the analysis of monoterpenes in indoor air by active sampling and thermal desorption-gas chromatography-mass spectrometry
Publication date :
15 September 2010
Journal title :
Talanta
ISSN :
0039-9140
eISSN :
1873-3573
Publisher :
Elsevier Science, Amsterdam, Netherlands
Volume :
82
Issue :
4
Pages :
1230-1239
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 18 October 2010

Statistics


Number of views
133 (8 by ULiège)
Number of downloads
2763 (30 by ULiège)

Scopus citations®
 
30
Scopus citations®
without self-citations
29
OpenCitations
 
25
OpenAlex citations
 
27

publications
0
supporting
0
mentioning
0
contrasting
0
Smart Citations
0
0
0
0
Citing PublicationsSupportingMentioningContrasting
View Citations

See how this article has been cited at scite.ai

scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

Bibliography


Similar publications



Sorry the service is unavailable at the moment. Please try again later.
Contact ORBi