[en] The aim of this study was to evaluate the practicality and accuracy of different attachment sites for the optodes of a pulse oximeter (measuring arterial haemoglobin oxygen saturation) in healthy cattle, and to assess the accuracy of pulse oximetry in diseased cattle with low haemoglobin oxygen saturation values caused by respiratory disease. The tait the nasal septum and the genital mucosa of females provided a continuous, stable and intense signal. The smallest bias, and no significant difference between measurements of arterial haemoglobin (mSp02) with the pulse oximeter and measurements of arterial haemoglobin oxygen saturation (SaOZ) with a blood gas analyser was obtained when the probe was attached to the tail. This site was used to evaluate the accuracy of pulse oximetry in animais with respiratory disease. There was a small bias between the measurements of Sa02 and mSp02, with a tendency for pulse oximetry to underestimate higher values and to overestimate lower values. The precision of pulse oximetry decreased substantially with the values for Sa02 <80 per cent, which lies outside the clinically relevant range
Disciplines :
Anatomy (cytology, histology, embryology...) & physiology Veterinary medicine & animal health
Author, co-author :
Coghe, Joost; Université de Liège - ULiège > Faculty of Veterinary Medicine > Large Clinical Sciences
Uystepruyst, Christophe; Université de Liège - ULiège > Faculty of veterinary Medicine > Large Clinical Sciences
Bureau, Fabrice ; Université de Liège - ULiège > Faculty of Veterinary Medicine > Large Clinical Sciences
Lekeux, Pierre ; Université de Liège - ULiège > Faculty of Veterinary Medicine > Large Clinical Sciences
Language :
English
Title :
Non-invasive assessment of arterial haemoglobin oxygen saturation in cattle by pulse oximetry
Publication date :
1999
Journal title :
Veterinary Record
ISSN :
0042-4900
eISSN :
2042-7670
Publisher :
British Veterinary Medicine, London, United Kingdom
ALTMAN, D. G. & BLAND, J. M. (1983) Measurement in medicine: the analysis of method comparison studies. Statistician 32, 307-317
BLAND, J. M. & ALTMAN, D. G. (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I, 307-310
CAULKETT, N. A., HAIGH, J. C. & CRIBB, F. H. (1996) Accuracy of non-invasive monitoring techniques in mobilised deer (Abstract). Veterinary Surgery 25, 180
CHAFFIN, M. K., MATTHEWS, N. S., COHEN, N. D. & CARTER, G. K. (1996) Evaluation of pulse oximetry in anaesthetised foals using multiple combinations of transducer type and transducer attachment site. Equine Veterinary Journal 28, 437-445
CLERBAUX, T. H., GUSTIN, P., DETRY, B., CAO, M. L. & FRANS, A. (1993) Comparative study of the oxyhaemoglobin dissociation curve for four mammals: man, dog, horse and cattle. Comparative Biochemistry and Physiology 106A, 687-694
DESMECHT, D. J-M., AMORY, H., LINDEN, A. S. & LEKEUX, P. (1996) Haemodynamic dysfunctions involved in experimental bovine pneumonic pasteurellosis. Journal of Experimental Animal Science 37, 173-189
DUNN, S. E., GODWIN, J., HOARE, R. J. T., KIRKLAND, P. D., WALKER, S. B., COVERDALE, O. R. & GIBSON, J. A. (1998) Diseases of feedlot cattle in Eastern Australia. Proceedings of the XXth World Buiatrics Congress, Sydney, Australia. pp 159-164
EMERY, J. R. (1987) Skin pigmentation as an influence on the accuracy of pulse oximetry. Journal of Perinatology 7, 329-330
ERHARDT, W., LENDL, C., HIPP, R., SCHINDELE, M. & BLÜMEL, G. (1989) Die Pulsoximetrie - ein nicht invasives Verfahren zur unmittelbaren und kontinuierlichen Überwachung von Sauerstoffsättigung und Pulsfrequenz - Vergleichsstudien zur Blutgasanalyse und zum Hämoreflektometer an Hund, Schwein und Schaf. Berliner und Münchener Teirärztliche Wochenschrift 102, 289-292
ESSELMONT, R. S., KOSSAIBATI, N. A. & REEVE-JOHNSON, L. (1998) The costs of respiratory disease in dairy heifer calves. Proceedings of the XXth World Buiatrics Congress, Sydney, Australia. pp 685-691
GRIFFIN, D. (1997) Economic impact associated with respiratory disease in beef cattle. Veterinary Clinics of North America: Food Animal Practice 13, 367-378
GROSENBAUGH, D. A. & MUIR, W. W. (1998) Pulse oximetry: a practical, efficient monitoring method. Veterinary Medicine 93, 60-66
HENDRICKS, J. C. & KING, L. G. (1994) Practicality, usefulness, and limits of pulse oximetry in critical small animal patients. Veterinary Emergency and Critical Care 3, 5-12
JACOBSON, J. D., MILLER, M. W., MATTHEWS, N. S., HARTSFIELD, S. M. & KNAUER, K. W. (1992) Evaluation of accuracy of pulse oximetry in dogs. American Journal of Veterinary Research 53, 537-540
JOHNSON, N. (1991) Monitoring fetal artertial oxygen saturation during labour (pulse oximetry). In Fetal and Neonatal Measurements. Eds J. G. Aarnoudse, H. N. Lafeber, H. W. Jongsma. Amsterdam, Excerpta Medica. pp 103-109
KELMAN, G. R. & NUNN, J-F. (1966) Nomograms for correction of blood PaO2, PCO2, pH and base excess for time and temperature. Journal of Applied Physiology 21, 1484-1490
LEKEUX, P., CLOSE, R. & ART, T. (1993) Pulmonary function testing in veterinary medicine. In Pulmonary Function in Healthy, Exercising and Diseased Animals. Ed P. Lekeux. Gand, VDT Publications, pp 15-43
LEMAIRE, M., ART, T., HENROTEAUX, M. & LEKEUX, P. (1991) Potentialités de l'application de l'oxymétrie à infrarouge basée sur la pulsation dans l'espèce canine. Annales de Médecine Vétérinaire 135, 203-212
MATTHEWS, N. S., HARTSFIELD, S. M., SANDERS, E. A., LIGHT, G. S. & SLATER, M. S. (1994) Evaluation of pulse oximetry in horses surgically treated for colic. Equine Veterinary Journal 26, 114-116
MOENS, Y., GOOTJES, P., LAGERWEIJ, E. & VAN DIJK, P. (1991) Überwachung der Sauerstoffsättigung von Pferden in Halothannarkose durch Pulsoximetrie am Septum Nasale. Berliner und Münchener Tierärztliche Wochenschrift 104, 357-360
MUYLLE, S., ANTHONE, P., SIMOENS, P. & LAUWERS, H. (1996) Preferential sites for arterial blood sampling in cattle. Veterinary Record 139, 86-88
NORTON, J. R. & MUNRO, J. (1996) Measuring arterial oxygen saturation in neonatal calves using pulse oximetry. Cattle Practice 4, 157-165
NORTON, J. R., JACKSON, P. G. G. & TAYLOR, P. M. (1998) Measurement of arterial oxygen-haemoglobin saturation in newborn lambs by pulse oximetry. Veterinary Record 142, 107-109
PETROIANU, G., MALEK, W. & BERGLER, W. F. (1994) Ideal oximetry site in pigs: tail or snout? Proceedings of the 5th International Congress of Veterinary Anaesthesia, Guelph, Canada, p 93
ROSSDALE, P. D. (1970) Some parameters of respiratory function in normal and abnormal newborn foals with special reference to levels of PaO2 during air and oxygen inhalation. Research in Veterinary Science 11, 270-276
SEVERINGHAUS, J. W. & ASTRUP, P. B. (1987) History of blood gas analysis. International Anesthesiology Clinics 24, 1-224
SEVERINGHAUS, J. W., ASTRUP, P. B. & MURRAY, J. F. (1998) Blood gas analysis and critical care medicine. American Journal of Respiratory and Clinical Care Medicine 157, S114-S122
SEVERINGHAUS, J. W. & KELLEHER, J. F. (1992) Recent developments in pulse oximetry. Anesthesiology 76, 1018-1038
TREMPER, K. K. & BARKER, S. J. (1989) Pulse oximetry. Anesthesiology 70, 98-108
UYSTEPRUYST, Ch., COGHE, J., BUREAU, F. & LEKEUX, P. (1999) Evaluation of accuracy of pulse-oximetry in newborn calves. Veterinary Journal (In press)
VEIT, H. P. & FARRELL, R. L. (1978) The anatomy and physiology of the bovine respiratory system relating to pulmonary disease. Cornell Veterinarian 68, 555-581
WHITEHAIR, K. J., WATNEY, G. C. G., LEITH, D. E. & DEBOWES, R. M. (1990) Pulse oximetry in horses. Veterinary Surgery 19, 243-248
Similar publications
Sorry the service is unavailable at the moment. Please try again later.
This website uses cookies to improve user experience. Read more
Save & Close
Accept all
Decline all
Show detailsHide details
Cookie declaration
About cookies
Strictly necessary
Performance
Strictly necessary cookies allow core website functionality such as user login and account management. The website cannot be used properly without strictly necessary cookies.
This cookie is used by Cookie-Script.com service to remember visitor cookie consent preferences. It is necessary for Cookie-Script.com cookie banner to work properly.
Performance cookies are used to see how visitors use the website, eg. analytics cookies. Those cookies cannot be used to directly identify a certain visitor.
Used to store the attribution information, the referrer initially used to visit the website
Cookies are small text files that are placed on your computer by websites that you visit. Websites use cookies to help users navigate efficiently and perform certain functions. Cookies that are required for the website to operate properly are allowed to be set without your permission. All other cookies need to be approved before they can be set in the browser.
You can change your consent to cookie usage at any time on our Privacy Policy page.