Customizable data acquisition system; Fluid infusion rate; Hemodynamic analysis; Invasive arterial pressure; Arterial pressures; Customizable; Data acquisition system; Data capture systems; Haemodynamics; Hemodynamic analyze; Intensive care; Civil and Structural Engineering; Biomedical Engineering; Instrumentation; Mechanical Engineering; Industrial and Manufacturing Engineering
Abstract :
[en] Patient specific cardiovascular system models have the potential to provide far greater insights into patient conditions than is currently possible in the intensive care unit (ICU). Access to haemodynamic data is imperative for the introduction and validation of these models for model-based care in the ICU. However, current bedside machines are proprietary systems and do not allow access to any data without potentially prohibitive cost, creating a barrier to research and advancements in bedside care. This work presents a novel, low-cost data capture system which allows real time capture of haemodynamic measurements, as well as fluid infusion rates in the ICU. The system consists of an Arduino controlled data capture unit, an adaptor for connecting to existing bedside pressure sensors, and a Python based application which displays and records data sent from the data capture system over a USB serial connection. The total system is highly customizable if needed, and costs a total of NZ$200.
Disciplines :
Engineering, computing & technology: Multidisciplinary, general & others
Author, co-author :
Cushway, James ; Université de Liège - ULiège > Aérospatiale et Mécanique (A&M) ; University of Canterbury, Department of Mechanical Engineering, 69 Creyke Road, Ilam, Christchurch 8041, New Zealand
Chase, J Geoffrey; University of Canterbury, Department of Mechanical Engineering, 69 Creyke Road, Ilam, Christchurch 8041, New Zealand
Desaive, Thomas ; Université de Liège - ULiège > Département d'astrophysique, géophysique et océanographie (AGO) > Thermodynamique des phénomènes irréversibles
Murphy, Liam; University of Canterbury, Department of Mechanical Engineering, 69 Creyke Road, Ilam, Christchurch 8041, New Zealand
Flett, Isaac L; University of Canterbury, Department of Mechanical Engineering, 69 Creyke Road, Ilam, Christchurch 8041, New Zealand
Shaw, Geoffrey M; Dept of Intensive Care, Christchurch Hospital, 2 Riccarton Avenue, Christchurch Central City, Christchurch 4710, New Zealand
Language :
English
Title :
A novel clinical data acquisition device: Towards real time cardiovascular modelling in the ICU.
H2020 - 872488 - DCPM - Digitalized Clone for Personalized Medicine
Funders :
EU - European Union
Funding text :
Funding: This work was supported by EU H2020 ERA Permed JTC2021 , \u201CPersonalised perfusion guided fluid therapy\u201D.Funding: The project was supported by EU H2020 R&I programme (MSCA-RISE-2019 call) under grant agreement #872488 \u2014 DCPM.
Kumar, R., Dubey, P., Zafer, A., Kumar, A., Yadav, S., Past, present and future of blood pressure measuring instruments and their calibration. Measurement, 172, 2021, 108845, 10.1016/j.measurement.2020.108845 [Online]. Available.
Cushway, J., Murphy, L., Chase, J.G., Shaw, G.M., Desaive, T., Physiological trend analysis of a novel cardio-pulmonary model during a preload reduction manoeuvre. Comput. Methods Programs Biomed., 220, 2022, 106819, 10.1016/j.cmpb.2022.106819 [Online]. Available.
Murphy, L., Davidson, S., Knopp, J.L., Chase, J.G., Zhou, T., Desaive, T., State analysis of total stressed blood volume and arterial elastance during induced sepsis. 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC, 2019, 2951–2954, 10.1109/EMBC.2019.8857518 [Online]. Available.
Pironet, A., Desaive, T., Geoffrey Chase, J., Morimont, P., Dauby, P.C., Model-based computation of total stressed blood volume from a preload reduction manoeuvre. Math. Biosci. 265 (2015), 28–39, 10.1016/j.mbs.2015.03.015 [Online]. Available.
Enev, R., Krastev, P., and, F.A., Prediction of fluid responsiveness: a review. Biotechnol. Biotechnol. Equip. 35:1 (2021), 1147–1155, 10.1080/13102818.2021.1960190 [Online]. Available.
Guarracino, F., Bertini, P., Pinsky, M.R., Cardiovascular determinants of resuscitation from sepsis and septic shock. Crit. Care, 23(1), 2019, 118, 10.1186/s13054-019-2414-9 [Online]. Available.
Pironet, A., Model-Based Prediction of the Response to Vascular Filling Therapy. 2016, University of Liège.
Davey, N.A., Murphy, L., Chase, J.G., Zhou, C., Model parameter identification as an index of fluid responsiveness. IFAC-PapersOnLine 58:24 (2024), 584–589, 10.1016/j.ifacol.2024.11.102 [Online]. Available. 12th IFAC Symposium on Biological and Medical Systems BMS 2024.
Smith, R., Rolfe, A., Cameron, C., Shaw, G.M., Chase, J.G., Pretty, C.G., Low cost circulatory pressure acquisition and fluid infusion rate measurement system for clinical research. HardwareX, 11, 2022, e00318, 10.1016/j.ohx.2022.e00318 [Online]. Available.