Quantitative capabilities of commercial CZT SPECT-CT cameras with 99mTc.
Seret, Alain; Bernard, Claire
2025In EJNMMI Physics, 12 (1), p. 44
Peer Reviewed verified by ORBi
 

Files


Full Text
s40658-025-00754-3.pdf
Publisher postprint (1.42 MB) Creative Commons License - Attribution, Non-Commercial, No Derivative
Download
Annexes
SupData_Rev2_R2.pdf
(3.22 MB) Creative Commons License - Attribution, Non-Commercial, No Derivative
http://creativecommons.org/licenses/by-nc-nd/4.0/.Seret and Bernard; EJNMMI Physics (2025) 12:44; https://doi.org/10.1186/s40658‐025‐00754‐3
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
CZT; Contrast; Quantification; Ring-shaped; SPECT; SPECT-CT
Abstract :
[en] [en] BACKGROUND: This study aimed to analyse the quantitative capabilities of cadmium-zinc-telluride (CZT)-based SPECT-CT cameras using 99mTc, comparable to the analysis performed a decade ago for the sodium iodide (NaI) SPECT-CT systems available on the market at that time. This survey assessed one dual-head (GE Discovery NM870 CZT) and two ring (GE Starguide, Spectrum Dynamics Veriton 200) CZT cameras, as well as a state-of-the-art NaI dual-head system (Siemens Intevo Bold) that served as reference. Attenuation and scatter correction accuracy was explored, contrast recovery for small cold and hot rods measured, as well as the quantification in a large uniform area using user-determined conversion factors. Tomography reconstruction was performed with the manufacturers' iterative algorithms that allowed for attenuation correction, scatter correction and resolution recovery. RESULTS: Using the NEMA NU-2 1994 dedicated phantom, attenuation and scatter corrections seemed to perform very well. Equally, the contrast recovery of cold rods seemed to be superior for the CZT systems. However, the contrast recovery for the hot rods was inferior to the NaI camera, whereas it was comparable without the scatter correction. Finally, a quantification error of less than 5% was shown to be reachable when using adequate user-determined conversion factors. For the NaI camera, all results were similar to those of the past study. CONCLUSIONS: Without scatter correction, the CZT SPECT systems showed contrast performance similar to the NaI camera. With scatter correction, this held true for cold objects but the contrast of hot objects was not significantly improved or was degraded depending on the system considered and the object size. Quantification errors of less than 5% were achievable. It is hoped that on-going developments at both manufacturers will improve the scatter correction accuracy.
Research Center/Unit :
GIGA CRC In vivo Imaging - ULiège
Disciplines :
Radiology, nuclear medicine & imaging
Author, co-author :
Seret, Alain  ;  Université de Liège - ULiège > Département de physique > Imagerie médicale expérimentale
Bernard, Claire ;  Médecine Nucléaire et Imagerie Oncologique, CHU Liège, Avenue de l'Hôpital, 1, 4000, Liège, Belgium
Language :
English
Title :
Quantitative capabilities of commercial CZT SPECT-CT cameras with 99mTc.
Publication date :
10 May 2025
Journal title :
EJNMMI Physics
eISSN :
2197-7364
Publisher :
Springer, Ch
Volume :
12
Issue :
1
Pages :
44
Peer reviewed :
Peer Reviewed verified by ORBi
Additional URL :
Available on ORBi :
since 02 June 2025

Statistics


Number of views
65 (3 by ULiège)
Number of downloads
21 (0 by ULiège)

Scopus citations®
 
1
Scopus citations®
without self-citations
0
OpenCitations
 
0
OpenAlex citations
 
3

Bibliography


Similar publications



Contact ORBi