[en] BACKGROUND: Pressure measuring systems using fluid-filled catheters can result in the recording of distorted pressure waveforms. It results in phase delay, overestimation of systolic and, to a lesser extent, of diastolic pressure. We designed and evaluated a method to correct this pressure waveform distortion using an appropriate transfer equation obtained from the dynamic response of the fluid-filled catheter. This transfer equation is based on the principle that a fluid-filled catheter recording system is considered as an underdamped dynamic system fully characterized by its natural frequency (omega n) and damping ratio (zeta). METHODS: Pressure waveforms, simultaneously recorded in vitro or in vivo by a fluid-filled catheter (Pc) and a micromanometer-tipped catheter (Pref), were used to validate the method. Dynamic response of the catheter used was obtained from a fastflush test. The corrected signal (Ppred) was obtained using omega n, zeta and the following transfer equation: d2Pc/dt2 + 2 omega n zeta dPc/dt + omega n 2Pc = C Ppred (t) After correction of Pc, Ppred was compared, using a linear regression, with Pref taken as reference. RESULTS: Our results showed that Ppred was fitted to Pref with excellent coefficient correlation (0.99). The mean error and the standard error of estimate were respectively -1.16 mmHg and 1.4 mmHg. CONCLUSION: This new method can convert the distorted pressure waveforms transmitted by any fluid-filled catheters into high-fidelity signals. It suppresses the phase delay and the over-estimation of systolic pressure induced by fluid-filled catheters.
Disciplines :
General & internal medicine
Author, co-author :
Lambermont, Bernard ; Centre Hospitalier Universitaire de Liège - CHU > Frais communs médecine
Gérard, Paul ; Université de Liège - ULiège > Département de mathématique > Statistique (aspects expérimentaux)
Detry, Olivier ; Centre Hospitalier Universitaire de Liège - CHU > Chirurgie abdominale- endocrinienne et de transplantation
Kolh, Philippe ; Université de Liège - ULiège > Département des Sciences biomédicales et précliniques > Service de biochimie et de physiologie humaines, normale et pathologique
Potty, P.
D'Orio, Vincenzo ; Université de Liège - ULiège > Département des sciences cliniques > Médecine d'urgence - bioch. et phys. hum. normales et path.
Marcelle, Roland ; Université de Liège - ULiège > Relations académiques et scientifiques (Médecine)
Language :
English
Title :
Correction of Pressure Waveforms Recorded by Fluid-Filled Catheter Recording Systems: A New Method Using a Transfer Equation
scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.
Glantz SA, Tyberg JV. Determination of frequency response from step response: application to fluid-filled catheters. Am J Physiol 1979: 5: H376-H378.
Kleinman B, Powell S, Kumar P, Gardner RM. The fast flush test measures the dynamic response of the entire blood pressure monitoring system. Anesthesiology 1992: 77: 1215-1220.
Shinozaki T, Deane RS, Mazuzan JE. The dynamic responses of liquid-filled catheter systems for direct measurements of blood pressure. Anesthesiology 1980: 53: 498-504.
Rothe CF, Kim KC. Measuring systolic arterial blood pressure. Crit Care Med 1980: 8: 683-689.
Bergel DH, Milnor WR. Pulmonary vascular impedance in the dog. Circ Res 1965: 16: 401-415.
Grant BJB, Paradowski LJ. Characterization of pulmonary arterial input impedance with lumped parameters models. Am J Physiol 1987: 252: H585-H593.
Hipkins SF, Rutten AJ, Runciman WB. Experimental analysis of catheter-manometer systems in vitro and in vivo. Anesthesiology 1989: 71: 893-906.
This website uses cookies to improve user experience. Read more
Save & Close
Accept all
Decline all
Show detailsHide details
Cookie declaration
About cookies
Strictly necessary
Performance
Strictly necessary cookies allow core website functionality such as user login and account management. The website cannot be used properly without strictly necessary cookies.
This cookie is used by Cookie-Script.com service to remember visitor cookie consent preferences. It is necessary for Cookie-Script.com cookie banner to work properly.
Performance cookies are used to see how visitors use the website, eg. analytics cookies. Those cookies cannot be used to directly identify a certain visitor.
Used to store the attribution information, the referrer initially used to visit the website
Cookies are small text files that are placed on your computer by websites that you visit. Websites use cookies to help users navigate efficiently and perform certain functions. Cookies that are required for the website to operate properly are allowed to be set without your permission. All other cookies need to be approved before they can be set in the browser.
You can change your consent to cookie usage at any time on our Privacy Policy page.