Essential information on surgical heart valve characteristics for optimal valve prosthesis selection: Expert consensus document from the European Association for Cardio-Thoracic Surgery (EACTS)-The Society of Thoracic Surgeons (STS)-American Association for Thoracic Surgery (AATS) Valve Labelling Task Force.
[en] Comprehensive information on the characteristics of surgical heart valves (SHVs) is essential for optimal valve selection. Such information is also important in assessing SHV function after valve replacement. Despite the existing regulatory framework for SHV sizing and labelling, this information is challenging to obtain in a uniform manner for various SHVs. To ensure that clinicians are adequately informed, the European Association for Cardio-Thoracic Surgery (EACTS), The Society of Thoracic Surgeons (STS) and American Association for Thoracic Surgery (AATS) set up a Task Force comprised of cardiac surgeons, cardiologists, engineers, regulatory bodies, representatives of the International Organization for Standardization and major valve manufacturers. Previously, the EACTS-STS-AATS Valve Labelling Task Force identified the most important problems around SHV sizing and labelling. This Expert Consensus Document formulates recommendations for providing SHV physical dimensions, intended implant position and hemodynamic performance in a transparent, uniform manner. Furthermore, the Task Force advocates for the introduction and use of a standardized chart to assess the probability of prosthesis-patient mismatch and calls valve manufacturers to provide essential information required for SHV choice on standardized Valve Charts, uniformly for all SHV models.
Disciplines :
Cardiovascular & respiratory systems
Author, co-author :
Durko, Andras P.
Pibarot, Philippe
Atluri, Pavan
Bapat, Vinayak
Cameron, Duke E.
Casselman, Filip P. A.
Chen, Edward P.
Dahle, Gry
Elefteriades, John A.
Lancellotti, Patrizio ; Université de Liège - ULiège > Département des sciences cliniques > Cardiologie - Pathologie spéciale et réhabilitation
Essential information on surgical heart valve characteristics for optimal valve prosthesis selection: Expert consensus document from the European Association for Cardio-Thoracic Surgery (EACTS)-The Society of Thoracic Surgeons (STS)-American Association for Thoracic Surgery (AATS) Valve Labelling Task Force.
The European Parliament and the Council of the European Union. Regulation (EU) 2017/745. Available at: http://eur-lex.europa.eu/legal-content/ENG/TXT/PDF/?uri=CELEX:32017R0745&from=EN. (Accessed 1 February 2020)
U.S. Food and Drug Administration. Premarket Approval (PMA) FDA. Available at: https://www.fda.gov/medical-devices/premarket-submissions/premarket-approval-pma. (Accessed 2 July 2020)
ISO. International. Standard, ISO 5840:2015. Cardiovascular Implants—Cardiac Valve Prostheses. International Organization for Standardization (ISO). Available at: www.iso.org, 2015. (Accessed 1 February 2020)
Durko, A.P., Head, S.J., Pibarot, P., Atluri, P., Bapat, V., Cameron, D.E., et al. Characteristics of surgical prosthetic heart valves and problems around labelling: a document from the European Association for Cardio-Thoracic Surgery (EACTS)—The Society of Thoracic Surgeons (STS)—American Association for Thoracic Surgery (AATS) Valve Labelling Task Force. Ann Thorac Surg 108 (2019), 292–303.
Frank, M., Ganzoni, G., Starck, C., Grünenfelder, J., Corti, R., Gruner, C., et al. Lack of accessible data on prosthetic heart valves. Int J Cardiovasc Imaging 32 (2016), 439–447.
Bapat, V.N., Attia, R., Thomas, M., Effect of valve design on the stent internal diameter of a bioprosthetic valve: a concept of true internal diameter and its implications for the valve-in-valve procedure. JACC Cardiovasc Interv 7 (2014), 115–127.
Ruzicka, D.J., Hettich, I., Hutter, A., Bleiziffer, S., Badiu, C.C., Bauernschmitt, R., et al. The complete supraannular concept: in vivo hemodynamics of bovine and porcine aortic bioprostheses. Circulation 120 (2009), S139–S145.
Cameron, D., Little things matter. J Thorac Cardiovasc Surg 149 (2015), 918–919.
Christakis, G.T., Buth, K.J., Goldman, B.S., Fremes, S.E., Rao, V., Cohen, G., et al. Inaccurate and misleadingvalve sizing: a proposed standard for valvesize nomenclature. Ann Thorac Surg 66 (1998), 1198–1203.
Walther, T., Falk, V., Weigl, C., Diegeler, A., Rauch, T., Autschbach, R., et al. Discrepancy of sizers for conventional and stentless aortic valve implants. J Heart Valve Dis 6 (1997), 145–148.
Bartels, C., Leyh, R.G., Matthias Bechtel, J.F., Joubert-Hubner, E., Sievers, H.H., Discrepancies between sizer and valve dimensions: implications for small aortic root. Ann Thorac Surg 65 (1998), 1631–1633.
Cochran, R.P., Kunzelman, K.S., Discrepancies between labeled and actual dimensions of prosthetic valves and sizers. J Card Surg 11 (1996), 318–324 [discussion: 25].
Ruzicka, D.J., Eichinger, W.B., Hettich, I.M., Bleiziffer, S., Bauernschmitt, R., Lange, R., Hemodynamic performance of the new St. Jude Medical Epic Supra porcine bioprosthesis in comparison to the Medtronic Mosaic on the basis of patient annulus diameter. J Heart Valve Dis 17 (2008), 426–433 [discussion: 34].
Lancellotti, P., Pibarot, P., Chambers, J., Edvardsen, T., Delgado, V., Dulgheru, R., et al. Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography and the Brazilian Department of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging 17 (2016), 589–590.
Wu, C., Saikrishnan, N., Chalekian, A.J., Fraser, R., Ieropoli, O., Retta, S.M., et al. In-vitro pulsatile flow testing of prosthetic heart valves: a round-Robin study by the ISO cardiac valves working group. Cardiovasc Eng Technol 10 (2019), 397–422.
Retta, S.M., Kepner, J., Marquez, S., Herman, B.A., S Shu MC, Grossman, L.W., In-vitro pulsatile flow measurement in prosthetic heart valves: an inter-laboratory comparison. J Heart Valve Dis 26 (2017), 72–80.
Rosenhek, R., Binder, T., Maurer, G., Baumgartner, H., Normal values for Doppler echocardiographic assessment of heart valve prostheses. J Am Soc Echocardiogr 16 (2003), 1116–1127.
Bavaria, J.E., Desai, N.D., Cheung, A., Petracek, M.R., Groh, M.A., Borger, M.A., et al. The St Jude Medical Trifecta aortic pericardial valve: results from a global, multicenter, prospective clinical study. J Thorac Cardiovasc Surg 147 (2014), 590–597.
Sabik, J.F. III, Rao, V., Lange, R., Kappetein, A.P., Dagenais, F., Labrousse, L., et al. One-year outcomes associated with a novel stented bovine peri-cardial aortic bioprosthesis. J Thorac Cardiovasc Surg 156 (2018), 1368–1377.e5.
Bleiziffer, S., Ali, A., Hettich, I.M., Akdere, D., Laubender, R.P., Ruzicka, D., et al. Impact of the indexed effective orifice area on mid-term cardiac-related mortality after aortic valve replacement. Heart 96 (2010), 865–871.
Pibarot, P., Dumesnil, J.G., Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol 36 (2000), 1131–1141.
Fallon, J.M., DeSimone, J.P., Brennan, J.M., O'Brien, S., Thibault, D.P., DiScipio, A.W., et al. The incidence and consequence of prosthesis-patient mismatch after surgical aortic valve replacement. Ann Thorac Surg 106 (2018), 14–22.
Head, S.J., Mokhles, M.M., Osnabrugge, R.L., Pibarot, P., Mack, M.J., Takkenberg, J.J., et al. The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27186 patients with 133141 patient-years. Eur Heart J 33 (2012), 1518–1529.
Pibarot, P., Magne, J., Leipsic, J., Côté, N., Blanke, P., Thourani, V.H., et al. Imaging for predicting and assessing prosthesis-patient mismatch after aortic valve replacement. JACC Cardiovasc Imaging 12 (2019), 149–162.
Bleiziffer, S., Eichinger, W.B., Hettich, I., Guenzinger, R., Ruzicka, D., Bauernschmitt, R., et al. Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method?. Heart 93 (2007), 615–620.
Cohen, R.G., Bourne, E.T., Industry-generated charts for the selection of stented aortic valve prostheses: clinical tool or marketing ploy?. Ann Thorac Surg 91 (2011), 1001–1012.
Verbraecken, J., Van de Heyning, P., De Backer, W., Van Gaal, L., Body surface area in normal-weight, overweight, and obese adults. A comparison study. Metabolism 55 (2006), 515–524.
Fatima, B., Mohananey, D., Khan, F.W., Jobanputra, Y., Tummala, R., Banerjee, K., et al. Durability data for bioprosthetic surgical aortic valve: a systematic review. JAMA Cardiol 4 (2019), 71–80.
Capodanno, D., Petronio, A.S., Prendergast, B., Eltchaninoff, H., Vahanian, A., Modine, T., et al. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: a consensus statement from the European Association of Percutaneous Cardiovascular Interventions (EAPCI) endorsed by the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur J Cardiothorac Surg 52 (2017), 408–417.
Dvir, D., Bourguignon, T., Otto, C.M., Hahn, R.T., Rosenhek, R., Webb, J.G., et al. Standardized definition of structural valve degeneration for surgical and transcatheter bioprosthetic aortic valves. Circulation 137 (2018), 388–399.