Article (Scientific journals)
Cutting balloon angioplasty for the prevention of restenosis: Results of the cutting balloon global randomized trial
Mauri, L.; Bonan, R.; Weiner, B. H. et al.
2002In American Journal of Cardiology, 90 (10), p. 1079-1083
Peer Reviewed verified by ORBi
 

Files


Full Text
cutting balloon.pdf
Publisher postprint (87.08 kB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Abstract :
[en] The cutting balloon (CB) is a specialized device designed to create discrete longitudinal incisions in the atherosclerotic target coronary segment during balloon inflation. Such controlled dilatation theoretically reduces the force needed to dilate an obstructive lesion compared with standard percutaneous transluminal coronary angioplasty (PTCA). We report a multicenter, randomized trial comparing the incidence of restenosis after CB angioplasty versus conventional balloon angioplasty in 1,238 patients. Six hundred seventeen patients were randomized to CB treatment, and 621 to PTCA. The mean reference vessel diameter was 2.86 +/- 0.49 mm, mean lesion length 8.9 +/- 4.3 mm, and prevalence of diabetes mellitus in patients was 13%. The primary end point, the 6-month binary angiographic restenosis rate, was 31.4% for CB and 30.4% for PTCA (p = 0.75). Acute procedural success, defined as the attainment of <50% diameter stenosis without in-hospital major adverse cardiac events, was 92.9% for CB and 94.7% for PTCA (p = 0.24). Freedom from target vessel revascularization was slightly higher in the CB arm (88.5% vs 84.6%, log-rank p = 0.04). Five coronary perforations occurred in the CB arm only (0.8% vs 0%, p = 0.03). At 270 days, rates of myocardial infarction, death, and total major adverse cardiac events for CB and PTCA were 4.7% versus 2.4% (p = 0.03), 1.3% versus 0.3% (p = 0.06), and 13.6% versus 15.1% (p = 0.34), respectively. In summary, the proposed mechanism of controlled dilatation did not reduce the rate of angiographic restenosis for the CB compared with conventional balloon angioplasty. CB angioplasty should be reserved for difficult lesions in which controlled dilatation is believed to provide a better acute result compared with balloon angioplasty alone. (C) 2002 by Excerpta Medica, Inc.
Disciplines :
Cardiovascular & respiratory systems
Author, co-author :
Mauri, L.
Bonan, R.
Weiner, B. H.
Legrand, Victor ;  Université de Liège - ULiège > Département des sciences cliniques > Cardiologie
Bassand, J. P.
Popma, J. J.
Niemyski, P.
Prpic, R.
Ho, K. K. L.
Chauhan, M. S.
Cutlip, D. E.
Bertrand, O. F.
Kuntz, R. E.
More authors (3 more) Less
Language :
English
Title :
Cutting balloon angioplasty for the prevention of restenosis: Results of the cutting balloon global randomized trial
Publication date :
15 November 2002
Journal title :
American Journal of Cardiology
ISSN :
0002-9149
eISSN :
1879-1913
Publisher :
Excerpta Medica Inc, New York, United States - New York
Volume :
90
Issue :
10
Pages :
1079-1083
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 31 May 2014

Statistics


Number of views
235 (1 by ULiège)
Number of downloads
996 (2 by ULiège)

Scopus citations®
 
134
Scopus citations®
without self-citations
118
OpenCitations
 
104

Bibliography


Similar publications



Contact ORBi