Modern három dimenziós konformális craniospinalis besugárzási technika.
Sebestyén, Z.; Kovács, P.; GULYBAN, Akos et al.
2011In Magyar Onkologia, 55 (3), p. 187-192
Peer Reviewed verified by ORBi
 

Files


Full Text
Sebestyén et al. - 2011 - Modern 3D conformal craniospinal radiotherapy planning method. - Magyar onkologia.pdf
Publisher postprint (624.81 kB)
Download

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Brain Neoplasms; Ependymoma; Humans; Imaging, Three-Dimensional; Medulloblastoma; Prone Position; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; Radiotherapy, Conformal; Treatment Outcome
Abstract :
[en] The main problem of craniospinal irradiation (CSI) is the matching of the fields. The use of a suitable technique is very important because matching of the fields is necessary to use for the optimal cancer irradiation of the long planning target volume (PTV). Since 2007, 8 patients have received CT-based, 3D-planned conformal CSI in our Institute. Patient immobilization was made in prone position in a vacuum bed, using skull and pelvis masks. Organ-at-risk (OAR) contours were made by radiographers. PTV was contoured by radiation oncologists. The prescribed dose to the PTV was 36 Gy with 1.8 Gy dose per fraction. In the planning process the following aspects were taken under consideration: all points of the PTV had to receive at least 95% of the prescribed dose (according to ICRU 50, 62); at junction field edges the overlapping parts were eliminated using a multisegmental technique, where the adjacent segment ends of the neighbouring fields were shifted two times 2 cm, so that the three equally weighted segments used in one field had 2-2 cm distance from each other. In the CSI planning the shape of the patient and so the length of the PTV has made a big emphasis on determining the number of field matching. Thus in some cases instead of two, only one field matching was sufficient - this could be achieved by increasing the source-to-skin distance (SSD) of the fields. The verification made with a solid-water phantom justified the precision of the field matching. The offset used at junction field edges in between one treatment facilitates the verification of field matching - and so the patient positioning. Thus the possibility of having overdosed regions could be reduced, which was very important from a radiation biological point of view.
Disciplines :
Radiology, nuclear medicine & imaging
Author, co-author :
Sebestyén, Z.;  Onkoterápiás Intézet, Pécsi Tudományegyetem, ÁOK KK, Pécs, Hungary.
Kovács, P.
GULYBAN, Akos ;  Centre Hospitalier Universitaire de Liège - CHU > Radiothérapie
Farkas, R.
Bellyei, S.
Szigeti, A.
Gallainé, D. F.
Mangel, L.
Language :
Hungarian
Title :
Modern három dimenziós konformális craniospinalis besugárzási technika.
Publication date :
2011
Journal title :
Magyar Onkologia
ISSN :
0025-0244
Publisher :
Magyar Onkologusok Tarsasaga, Budapest, Hungary
Volume :
55
Issue :
3
Pages :
187-192
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 20 June 2013

Statistics


Number of views
55 (0 by ULiège)
Number of downloads
145 (0 by ULiège)

Scopus citations®
 
2
Scopus citations®
without self-citations
0

Bibliography


Similar publications



Contact ORBi