Keywords :
Matching; Mechanical ventilation; NAVA; Patient-ventilator interaction; Variability
Abstract :
[en] Neurally adjusted ventilatory assist (NAVA) is a ventilation assist mode that delivers pressure in proportionality to electrical activity of the diaphragm (Eadi). Compared to pressure support ventilation (PS), it improves patient-ventilator synchrony and should allow a better expression of patient's intrinsic respiratory variability. We hypothesize that NAVA provides better matching in ventilator tidal volume (Vt) to patients inspiratory demand. 22 patients with acute respiratory failure, ventilated with PS were included in the study. A comparative study was carried out between PS and NAVA, with NAVA gain ensuring the same peak airway pressure as PS. Robust coefficients of variation (CVR) for Eadi and Vt were compared for each mode. The integral of Eadi (sh{phonetic}Eadi) was used to represent patient's inspiratory demand. To evaluate tidal volume and patient's demand matching, Range90 = 5-95 % range of the Vt/sh{phonetic}Eadi ratio was calculated, to normalize and compare differences in demand within and between patients and modes. In this study, peak Eadi and sh{phonetic}Eadi are correlated with median correlation of coefficients, R > 0.95. Median sh{phonetic}Eadi, Vt, neural inspiratory time (Ti_ <br /> Neural), inspiratory time (Ti) and peak inspiratory pressure (PIP) were similar in PS and NAVA. However, it was found that individual patients have higher or smaller sh{phonetic}Eadi, Vt, Ti_ <br /> Neural, Ti and PIP. CVR analysis showed greater Vt variability for NAVA (p < 0.005). Range90 was lower for NAVA than PS for 21 of 22 patients. NAVA provided better matching of Vt to sh{phonetic}Eadi for 21 of 22 patients, and provided greater variability Vt. These results were achieved regardless of differences in ventilatory demand (Eadi) between patients and modes. © 2012 Springer Science+Business Media New York.
Disciplines :
Anesthesia & intensive care
General & internal medicine
Author, co-author :
Moorhead, K. T.;  Cardiovascular Research Center, University of Liege, Liege, Belgium
Piquilloud, L.;  Intensive Care and Burn Unit, University Hospital, Lausanne, Switzerland
LAMBERMONT, Bernard  ;  Centre Hospitalier Universitaire de Liège - CHU > Frais communs médecine
Roeseler, J.;  Intensive Care Unit, Cliniques Universitaires St-Luc, Brussels, Belgium
Chiew, Y. S.;  Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand
Chase, J. G.;  Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand
Revelly, J.-P.;  Intensive Care and Burn Unit, University Hospital, Lausanne, Switzerland
Bialais, E.;  Intensive Care Unit, Cliniques Universitaires St-Luc, Brussels, Belgium
Tassaux, D.;  Intensive Care Unit, University Hospital, Geneva, Switzerland
Laterre, P.-F.;  Intensive Care Unit, Cliniques Universitaires St-Luc, Brussels, Belgium
Jolliet, P.;  Intensive Care and Burn Unit, University Hospital, Lausanne, Switzerland
Sottiaux, T.;  La Clinique Notre Dame de Grâce, Gosselies, Belgium
Desaive, Thomas  ;  Université de Liège - ULiège > Département d'astrophys., géophysique et océanographie (AGO) > Thermodynamique des phénomènes irréversibles
More authors (3 more) Less
Language :
English
Title :
NAVA enhances tidal volume and diaphragmatic electro-myographic activity matching: a Range90 analysis of supply and demand
Publication date :
2013
Journal title :
Journal of Clinical Monitoring and Computing
ISSN :
1387-1307
eISSN :
1573-2614
Publisher :
Kluwer Academic Publishers, Netherlands
Volume :
27(1)
Pages :
61-70
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 05 October 2012

Statistics


Number of views
145 (13 by ULiège)
Number of downloads
626 (13 by ULiège)

Scopus citations®
 
21
Scopus citations®
without self-citations
5
OpenCitations
 
19

Bibliography


Similar publications