Article (Scientific journals)
Comparison of several strategies for the deployment of a multivariate regression model on several handheld NIR instruments. Application to the quality control of medicines
Ciza Hamuli, Patient; Sacre, Pierre-Yves; Waffo Tchounga, Christelle Ange et al.
2022In Journal of Pharmaceutical and Biomedical Analysis, 215, p. 114755
Peer Reviewed verified by ORBi
 

Files


Full Text
1-s2.0-S0731708522001765-main.pdf
Publisher postprint (2.4 MB)
Request a copy
Annexes
supplementary materials_PYS.docx
(1.87 MB)
Request a copy

All documents in ORBi are protected by a user license.

Send to



Details



Keywords :
Metformin; NIR spectroscopy; Handheld transmission spectrophotometer; Direct Standardization; Spectral Space Transformation; Slope/Bias Correction; Global modelling
Abstract :
[en] Chemometrics applied to spectroscopic measurements such as near-infrared are gaining more and more importance for quality control of pharmaceutical products. Handheld near-infrared devices show great promise as a medicines quality screening technique for post-marketing surveillance. These devices are able to detect substandard and falsified medicines in pharmaceutical supply chains and enable rapid action before these medicines reach patients. The instrumental and environmental changes, expected or not, can adversely affect the analytical performances of prediction models developed for routine applications. Based on a previous study, PLS prediction models were developed and validated on three similar handheld NIR transmission spectrophotometers of the same model and from same company. These models have shown to be effective in analyzing metformin tablet samples, but significant spectral differences between handheld systems complicated their deployment for routine analysis. In this study, different strategies have been applied and compared to correct the instrumental variations, including global modelling (GM) and calibration transfer methods (Direct Standardization, DS; Spectral Space Transformation, SST and Slope/Bias correction, SBC), considering the RMSEP and the accuracy profile as assessment criteria. The transfer methods showed good capabilities to maintain the predictive performances comparable to that of the global modelling approach, except for a remaining slight bias. This approach is interesting since very few standardization samples are required to develop an adequate transfer model. GM, SST and SBC were able to correct/handle drifts in the spectral responses of different handheld instruments and thus may help to avoid the need for a long, laborious, and costly full recalibration process due to inter-instrument variations.
Research center :
CIRM - Centre Interdisciplinaire de Recherche sur le Médicament - ULiège
Disciplines :
Pharmacy, pharmacology & toxicology
Author, co-author :
Ciza Hamuli, Patient  ;  Université de Liège - ULiège > Faculté de Médecine > Doct. sc. bioméd. & pharma. (paysage) ; University of Kinshasa, Faculty of Pharmaceutical Sciences, LACOMEDA
Sacre, Pierre-Yves  ;  Université de Liège - ULiège > Unités de recherche interfacultaires > Centre Interdisciplinaire de Recherche sur le Médicament (CIRM)
Waffo Tchounga, Christelle Ange  ;  Université de Liège - ULiège > Unités de recherche interfacultaires > Centre Interdisciplinaire de Recherche sur le Médicament (CIRM) ; University of Yaoundé I, Faculty of Medicine and Biomedical Sciences
Kimbeni, T. M.;  University of Kinshasa, Faculty of Pharmaceutical Sciences, LACOMEDA
Masereel, Bernard ;  Université de Liège - ULiège > Département de pharmacie
Hubert, Philippe  ;  Université de Liège - ULiège > Unités de recherche interfacultaires > Centre Interdisciplinaire de Recherche sur le Médicament (CIRM)
Ziemons, Eric  ;  Université de Liège - ULiège > Unités de recherche interfacultaires > Centre Interdisciplinaire de Recherche sur le Médicament (CIRM)
Marini Djang'Eing'A, Roland ;  Université de Liège - ULiège > Unités de recherche interfacultaires > Centre Interdisciplinaire de Recherche sur le Médicament (CIRM)
Language :
English
Title :
Comparison of several strategies for the deployment of a multivariate regression model on several handheld NIR instruments. Application to the quality control of medicines
Publication date :
July 2022
Journal title :
Journal of Pharmaceutical and Biomedical Analysis
ISSN :
0731-7085
eISSN :
1873-264X
Publisher :
Elsevier, Amsterdam, Netherlands
Volume :
215
Pages :
114755
Peer reviewed :
Peer Reviewed verified by ORBi
Funders :
ARES - Académie de Recherche et d'Enseignement Supérieur [BE]
F.R.S.-FNRS - Fonds de la Recherche Scientifique [BE]
Funding number :
PFS-2016; EOS Chimic
Available on ORBi :
since 04 April 2022

Statistics


Number of views
152 (17 by ULiège)
Number of downloads
12 (11 by ULiège)

Scopus citations®
 
8
Scopus citations®
without self-citations
4
OpenCitations
 
3

Bibliography


Similar publications



Contact ORBi